Objective evaluation of "unmeasurable materials"
The DigiSpec® series image spectrophotometer is a groundbreaking color measurement instrument that can be used in virtually all areas of color measurement. The image spectrophotometer can acquire color and spectral data for each point in the entire image per minute, with an area as small as 0.0016mm² (0.04mm*0.04mm), between square inches.
The image spectrophotometer uses an LED or xenon light source as the illumination source, which ensures that the traceability data is consistent with the previous generation of conventional spectrophotometers.
DigiSpec® Series image spectrophotometer can be applied to almost all color measurement fields
Textile industry: yarn, printing, lace, zipper, embroidery, etc
Construction industry: rock color paint, real stone paint, etc
Printing industry: wood grain paper, wallpaper, etc
Restoration of Cultural Relics: murals, calligraphy and painting, ceramics, etc
Others: buttons, jewelry, etc
DigiSpec ® The color and spectral data of each point in the whole image can be obtained. In the area of 0.04mm * 0.04mm, the minimum measurement can be 0.0016mm ²
DigiSpec® It is helpful for the consistency of color evaluation in the whole production process
The maximum value of measurement repeatability deab ≤ 0.01, whether measuring white standard plate or black standard plate digiscpec & reg; It can achieve the maximum measurement repeatability, deab ≤ 0.01 ※, and put forward a new standard of benchmark significance for desktop spectrophotometer
The perfect calibration structure and algorithm provide completely reliable short-term measurement stability and long-term measurement stability
※The surface color of different materials is greatly affected by temperature. Deab ≤ 0.01 is an extremely precise measurement stability condition. In the inspection of digiscpec & reg; When measuring repeatability, be sure to ensure the surface temperature stability of the measured object
Using LED or xenon light source as lighting source can ensure the consistency of traceability data with the previous generation of traditional spectrophotometer
There are 20 + standard light sources and 30 + measurement indicators. SCI / SCE measurement conditions cover all the functions of the previous generation of spectrophotometer, and can provide the measurement parameters and conditions of the traditional spectrophotometer
Model | DS-1000 | DS-1050 | DS-1100 |
Geometry | Dual beam D / 8 ° , SCI / SCE | ||
Light source | Full spectrum LED | High precision like sunlight full spectrum LED | High precision analog fluorescent xenon lamp light source |
Sphere diameter | 152mm/6in | ||
Wavelength range |
400nm-700nm |
400nm-1000nm |
|
Reporting interval | 10nm | 2.5nm | |
Photometric range | 0-200%, resolution 0.01% | ||
Repeated measurement of 30 times of double flash |
ΔE*ab≤0.03(Maximum) | ΔE*ab≤0.01(Maximum) | |
Inter-Instrument consistency for reflectance measurement |
0.4 | 0.25 | |
Aperture plate | Lav (30mm lighting, 25mm measurement), the user can set the measurement area and customize the caliber of any size | ||
Test conditions |
Compliance with standards:CIE No.15、GB/T 3978、GB 2893GB/T 18833、ISO7724/1、DIN5033 Teil7、 JIS Z8722 condition C、ASTM E1164、ASTM-D1003-07 |
||
Sensor | CMOS array sensor |
Silicon based metal oxide image sensor |
|
Spectroscopic modes | grating | Ultra high precision holographic transmission volume grating | |
Image resolution | 300dpi | 500dpi(high resolution) | 600dpi(Ultra high resolution) |
Minimum measurement area | better than 0.01mm² (0.1*0.1mm) | better than 0.004mm² (0.06*0.06mm) | better than 0.0016mm² (0.04*0.04mm) |
Observer | 2 ° standard observer and 10 ° standard observer | ||
light source | A,C,D50,D55,D65,D75,F1,F2,F3,F4,F5,F6,F7,F8,F9,F10,F11,F12,CWF,U30,DLF,NBF,TL83,TL84 | ||
Color space | L*a*b,L*C*h,Hunter Lab,Yxy,XYZ | ||
Color index |
WI(ASTM E313-00,ASTM E313-73,CIE/ISO,AATCC,Hunter,Taube,Berger Stensby), YI(ASTM D1925,ASTM E313-00,ASTM E313-73),Tint(ASTM E313-00),Isochromatic index milm, color fastness, Color fastness, ISO brightness, R457, a density, t density, e density, m density, hiding power, strength |
||
Colour-difference formula | ΔE*ab,ΔE*CH,ΔE*uv,ΔE*cmc,ΔE*94,ΔE*00,ΔEab(Hunter),555 hue classification | ||
Test time | Less than 8 seconds | Less than 5 seconds | |
Working temperature and humidity | 5 ~ 40 ° C, no condensation below 80% relative humidity (at 35 ° C) | ||
Storage temperature and humidity | -20 ~ 45 ° C, no condensation below 80% relative humidity (at 35 ° C) | ||
Accessories | Power adapter, data cable, whiteboard | ||
Interface | USB 3.0 |
※ The measuring caliber is 25 * 25MM. After the instrument is calibrated, measure the BCRA white correction plate 30 times at an interval of 10 seconds
※※ Average value of measurement results of 12 BCRA ceramic standard color plates