Satisfy color measurement anytime, anywhere
DS-700D spectrophotometer has good interstage difference and repeatability, and the measurement data is accurate and reliable. It adopts 400-700nm full wavelength LED light source and is equipped with 3 kinds of measurement apertures, which can meet the measurement of various samples. It not only supports connecting to computer to process data, but also supports Android, ios and WeChat applet, and data can be stored anywhere and anytime.
Repeatability ΔE*ab≤0.025, difference between units≤0.25, accurate and reliable data, ensuring the consistency of measurement data from multiple devices for color matching and accurate color transfer
Dual-column high-precision CMOS array design monitors the energy fluctuation of the light source while measuring the sample signal to reduce interference during measurement and obtain higher measurement stability. Large-area dual-array sensor provides higher spectral response sensitivity, ensuring high standards of instrument measurement speed, accuracy, stability and consistency of inter-stage differences
The instrument is placed on an intelligent calibration base, and the instrument can be automatically calibrated by the whiteboard on the base without human intervention. It changes the status quo of traditional instruments that must be calibrated manually on the white board. The white board made of ZrO2 with reflectivity R%>90% is used as the calibration standard, which ensures good mechanical strength and weather resistance, and no scratches and discoloration will occur in long-term use
The instrument uses a full-band balanced LED light source and UV light source as the illumination source, ensuring sufficient spectral distribution in the visible and UV bands, and materials containing fluorescent components can also be measured
The combination of a grating and an array sensor made by an innovative MEMS process allows for more accurate color measurement based on a 10 nm spectral resolution
Equipped with three measuring calipers, which can be replaced at any time without tools to meet the needs of various sample measurements
Provide a variety of measurement parameters and evaluation light sources in accordance with international standards Flexible to add various light sources and measurement indexes through mobile terminal or PC terminal program
Support Android, IOS, WeChat applet, Windows program
Camera view positioning, can clearly observe the measured area
Model | DS-700D | ||
Geometry
|
d/8,SCI+SCE | ||
Sensor |
Dual optical path high precise CMOS array sensor
|
||
Spectroscopic method | Grating | ||
Sphere size | Φ40mm | ||
Wavelength range | 400-700nm | ||
wavelength interval | 10nm | ||
Reflectance measurement range | 0-200%,Resolution:0.01% | ||
Lighting source | Full-band balanced LED light source + UV | ||
Measure time | About 1.5 s | ||
Illumination Area Aperture | 3 Apertures(MAV:Φ8mm/Φ11mm;SAV:Φ4mm/Φ6mm;MINI:1*3mm) | ||
Repeatability※ | ΔE*ab≤ 0.025 | ||
Inter-instrument agreement | ΔE*ab≤0.25 | ||
Language | Chinese, English | ||
Display Resolution | 0.01 | ||
Observer | 2°,10° | ||
Illuminants | A,B,C,D50,D55,D65,D75,F1,F2,F3,F4,F5,F6,F7,F8,F9,F10,F11,F12,CWF,U30,U35,DLF,NBF,TL83,TL84,ID50, ID65,LED-B1,LED-B2,LED-B3,LED-B4,LED-B5,LED-BH1,LED-RGB1,LED-V1,LED-V2 |
||
Color Space / Indices |
Reflectance, CIE-Lab, CIE-LCh, HunterLab, CIE Luv, XYZ, Yxy, RGB, Color difference(ΔE*ab, ΔE*cmc, ΔE*94,ΔE*00),WI(ASTM E313-00,ASTM E313-73,CIE/ISO, AATCC, Hunter, Taube Berger Stensby), YI(ASTM D1925,ASTM E313-00,ASTM E313-73), Blackness(My,dM),Color Fastness
|
||
Storage | APP mass storage | ||
Interface | USB,Bluetooth | ||
Battery | Rechargeable, 8000 times continuous tests, 3.7V/3000mAh | ||
Camera to see the measurement area | Yes | ||
Calibration | Auto Calibration | ||
Software | Andriod,IOS,Windows | ||
Accuracy Guarantee |
Metrology Level I
|
||
Standards |
Conform to CIE No.15, GB/T 3978, GB 2893, GB/T 18833, ISO7724-1, ASTM E1164, DIN5033 Teil7
|
||
Screen | IPS full color screen 3.5 inches | ||
Light source lifetime | 1 million times in 10 years | ||
Data storage | 100 targets, 10000 samples |
※ After the instrument is calibrated, measure the white calibration plate 30 times at 5-second intervals and measure the standard deviation of the results with the MAV caliber
※※ Average value of 12 swatches of BCRA series II