UV/Visible spectrophotometer with double-beam optical system, designed for applications including general research, pharmaceutical, biochemical and clinical laboratory use, new material development and quality control.
Delivery information: Supplied with four optical glass cells, two quartz cells, dust cover, USB cable, USB (8 GB) stick, UV-Vis Analyst software for PC control, cell holders for sample and reference, power cables with UK, CH and EU plugs.
SPECIFICATIONS:
Display | LCD, 320×240 P | LCD, 320×240 P | |
Optical design | Double beam | Double beam | |
Absorbance range | −0,3...+3,0 A | −0,3...+3,0 A | |
Absorbance accuracy | ±0,002 A @ 1,0 A | ±0,002 A @ 1,0 A | |
Photometric accuracy | ≤0,3% T | ≤0,3% T | |
Photometric range | 0 - 200% T; 0 - 9999 conc. | 0 - 200% T; 0 - 9999 conc. | |
Photometric stability | 0,001 A/h @ 500 nm | 0,001 A/h @ 500 nm | |
Wavelength accuracy | ±0,3 nm | ±0,3 nm | |
Wavelength range | 190 - 1100 nm | 190 - 1100 nm | |
Electrical | 110/220 V, 50/60 Hz | 110/220 V, 50/60 Hz | |
Interfaces | Parallel printer | Parallel printer | |
Languages | English, French, German, Spanish | English, French, German, Spanish | |
Light source | Tungsten/deuterium | Tungsten | |
Memory | 200 results and 200 curves | 200 results and 200 curves | |
Output | USB | USB | |
Scanning speed | 100 - 3000 nm/min Intervals: 0,1; 0,2; 0,5; 1,0; 5,0 nm | 100 - 3000 nm/min Intervals: 0,1; 0,2; 0,5; 1,0; 5,0 nm | |
Stray light | ≤0,05% T @ 220 and 360 nm | ≤0,05% T @ 220 and 360 nm | |
W×D×H | 590×420×260 mm | 590×420×260 mm | |
Weight | 26 kg | 26 kg |