TECHNICAL SPECIFICATIONS
MODEL NO. |
VH-5L VH-5DL |
VH-10L VH-10DL |
VH-30L VH-30DL |
VH-50L VH-50DL |
Vickers scales |
HV0.3, HV0.5, HV1, HV2, HV3, HV5 |
HV0.5, HV1, HV2, HV3, HV5, HV10 |
HV1, HV2.5, HV5, HV10, HV20, HV30 |
HV1, HV5, HV10, HV20, HV30, HV50 |
Test force selection |
Dial |
|||
Accuracy |
Conforms to EN-ISO 6507, ASTM E384,JIS B7734 |
|||
Load control |
Automatic (loading/dwell/unloading) |
|||
Load duration |
5 to 60 sec (5 sec increments) |
|||
Objectives |
10x, 20x |
5x, 10x |
||
Eyepiece magnification |
15x with digital encoder |
|||
Total magnification |
150x, 300x, |
75x, 150x |
||
Measuring range |
HV5 - HV3000 |
|||
Maximum specimen height |
160mm |
|||
Maximum specimen depth |
135mm from the centre line |
|||
Display |
Menu operated keypad with large LCD-screen |
|||
Hardness value |
5-digit |
|||
Diagonal length |
4-digit (D1, D2) |
|||
Load duration |
Dwell time (sec) |
|||
Conversion |
Rockwell, Superficial Rockwell, Brinell, Shore, Tensile |
|||
Statistics |
No., average, S. dev., R, real time after each test |
|||
Optical path |
2-way switchable : eyepiece/camera |
|||
Power supply |
220V, 50Hz |
|||
Dimensions |
220mm x 540mm x 650mm |
|||
Weight |
38kg |
STANDARD DELIVERY ● Main unit ● Diamond indentor Vickers ● Objectives 10x, 20x or 5x, 10x ● Digital eyepiece 15x ● Flat anvil ø 60mm ● Small V-type anvil ● Large V-type anvil ● Built-in printer ● RS232 data output ● 4 adjustable feet ● Level gauge ● Vickers test plate ● Spare halogen lamp ● Installation & users manual |
OPTIONAL ACCESSORIES ● Precision vice ● XY-stage with micrometers ● Indentors ● Test blocks ● CCD Video measurement system optional |
VICKERS HARDNESS VISION MEASUREMENT iVPAD
1.Features:
iVPad was developed by Windows operation system, it integrated high resolution CMOS camera and iV2.0
vickers software, iVPad can be worked with any vickers hardness testers, replace the traditional eyepiece,
directly measure the indentation, mainly used to upgrade existed or new vickers hardness testers.
There is no need to aim indentation and reading dialog line, and can avoid human error when measure
dialog line, largely improve accuracy and efficiency.
Users can export Word/Excel report, and feedback the result to their suppliers and customers, improve
quality control capacity.
2.Technical Data:
Operation System |
Win10 |
Size of Tablet |
10.6” |
USB Connector |
Double USB Connector |
CPU |
Intel I3 |
RAM |
2GB |
Hard Disc |
32GB |
Camera |
1.3MP Pixel, 1/2”CMOS Color Camera |
Vickers Software |
iV2.0 Tablet Vickers Software |
Application |
Any Vickers Hardness Testers |
Vickers Measurement Software iV2.0
1. Introduction:
Our new generation vickers/knoop measurement software with friendly operation interface, apply USB
camera to fast capture video and images, supports case depth inspection by analyzing the indentation
image.
In traditional way, operators judge the size of the indentation through the microscopes, due to vision
errors, different operators obtain different measurement results for same indentation, while by software,
then indentation is enlarged through high resolution lens, and the size of indentation can be automatical-
ly obtained by image processing algorithm, largely improve the accuracy and efficiency.
2. Features:
● One-Key operation, easy and fast!
● Automatic image process technology, high accuracy and stable result.
● curve chart to display hardness results.
● Support English, Russian and Chinese operation language.
● Standard measurement parameters and user-defined parameter for selection.
● Supports hardness conversion.
● Support report generation by Word/Excel.
3. Interface of Software:
Parameter Setting: Hardness Conversion:
More detail about LUX & BRILLIANCE TECHNOLOGIES SDN. BHD.