CNX 3000 Wireless Multimeter

For all specifications: accuracy is specified for one year after calibration, at operating temperatures of 18 °C to 28 °C, with relative humidity at 0 % to 90 %. Accuracy specifications take the form of ± ([ % of Reading ] + [ Number of least significant digits ]).

Detailed Specifications
AC Voltage
Range ¹ Resolution Accuracy ² ³
45 Hz to 500 Hz 500 Hz to 1 kHz
600.0 mV 0.1 mV 1.0 % + 3 2.0 % + 3
6.000 V 0.001 V
60.00 V 0.01 V
600.0 V 0.1 V
1000 V 1 V
¹ All AC voltage ranges are specified from 1 % of range to 100 % of range. 
² Crest factor of ≤ 3 at full scale p to 500 V, decreasing linearly to crest factor < 1.5 at 1000 V. 
³ For non-sinusoidal waveforms, add –(2 % of reading + 2 % full scale) typical, for crest factor up to 3.

DC Voltage, Continuity, Resistance, Diode Test and Capacitance
Function Range Resolution Accuracy
mV 600.0 mV 0.1 mV 0.09 % + 2
V 6.000 V 0.001 V 0.09 % + 2
60.00 V 0.01 V
600.0 V 0.1 V
1000 V 1 V 0.15 % + 2
600 Ω 1 Ω Meter beeps at < 25 Ω, beeper detects opens or shorts of 250 μs or longer.
600.0 Ω 0.1 Ω 0.5 % + 2
6.000 kΩ 0.001 kΩ 0.5 % + 1
60.00 kΩ 0.01 kΩ
60.00 kΩ 0.1 kΩ
600.0 kΩ 0.001 MΩ
50.00 MΩ 0.01 MΩ 1.5 % + 3
Diode test 2.000 V 0.001 V 1 % + 2
µF 1000 nF 1 nF 1.2 % + 2
10.00 μF 0.01 μF
100.0 μF 0.1 μF
9999 μF¹ 1 μF 10 %
¹ In the 9999 μF range for measurements to 1000 μF, the measurement accuracy is 1.2 % + 2.

AC and DC Current
Function Range ¹ Resolution Accuracy
mA AC 
(45 Hz to 1 kHz)
60.00 mA 0.01 mA 1.5 % + 3
400.0 mA ³ 0.1 mA
mA DC ² 60.00 mA 0.01 mA 0.5 % + 3
400.0 mA 0.1 mA
¹ All AC current ranges are specified from 5 % of range to 100 % of range. 
² Input burden voltage (typical): 400 mA input 2 mV/mA. 
³ 400.0 mA accuracy specified up to 600 mA overload.

Frequency
Range Resolution Accuracy ¹
99.99 Hz 0.01 Hz 0.1 % + 1
999.9 Hz 0.1 Hz
9.999 kHz 0.001 kHz
99.99 kHz 0.01 kHz
¹ Frequency is specified up to 99.99 kHz in volts and up to 10 kHz in amps.

Input Characteristics
Function Overload 
Protection
Input Impedance 
(nominal)
Common Mode Rejection Ratio (1 kΩ unbalance) Normal Mode Rejection
1100 V RMS > 10 MΩ < 100 pF > 120 dB at DC, 50 Hz or 60 Hz > 60 dB at 50 Hz or 60 Hz
1100 V RMS > 10 MΩ < 100 pF > 60 dB, DC to 60 Hz  
1100 V RMS > 10 MΩ < 100 pF > 120 dB at DC, 50 Hz or 60 Hz > 60 dB at 50 Hz or 60 Hz
Open circuit test voltage Full scale Typical short circuit current
To 6 MΩ 50 MΩ
1100 V RMS < 2.7 V DC < 0.7 V DC < 0.9 V DC < 350 mA
1100 V RMS < 2.7 V DC 2.000 V DC < 1.1 mA

Function Overload Protection Overload
mA Fused, 44/100 A, 1000 V FAST Fuse 600 mA overload for 2 minutes maximum, 10 minutes rest minimum

MIN/MAX Recording
Function Accuracy
DC functions The specified accuracy of the measurement function ± 12 counts for changes > 350 mS in duration.
AC functions The specified accuracy of the measurement function ± 40 counts for changes > 900 mS in duration.

General Specifications
Maximum voltage between any terminal and earth ground 1000 V DC or AC RMS
Ω fuse protection from A inputs 0.44 A (44/100 A, 440 mA), 1000 V FAST Fuse, Fluke specified part only
Display (LCD) Update rate: 4/sec
Volts, amps, ohms: 6000 counts
Frequency: 10,000 counts
Capacitance: 1,000 counts
Battery type Three AA Alkaline batteries, NEDA 15A IEC LR6
Battery life 300 hours minimum
Temperature Operating: -10 °C to 50 °C Storage: -40 °C to 60 °C
Relative humidity 0 % to 90 % (0 °C to 35 °C)
0 % to 75 % (35 °C to 40 °C)
0 % to 45 % (40 °C to 50 °C)
Altitude Operating: 2,000 m 
Storage: 12,000 m
Temperature coefficient 0.1 X (specified accuracy) /°C (<18 °C or >28 °C)
Wireless frequency 2.4 GHz ISM Band 20 meter range
Size (H x W x L) 4.75 cm x 9.3 cm x 20.7 cm
(1.87 in x 3.68 in x 8.14 in)
Weight 340 g (12 oz)
Safety standards US ANSI: ANSI/ISA 61010-1 / (82.02.01): 3rd edition
CSA: CAN/CSA-C22.2 No 61010-1-12: 3rd edition
CE European: IEC/EN 61010-1:2010
Electromagnetic compatibility 
EMI, RFI, EMC, RF
EN 61326-1:2006, EN 61326-2-2:2006. ETSI EN 300 328 V1.7.1:2006, ETSI EN 300 489 
V1.8.1:2008, FCC Part 15 Subpart C Sections 15.207, 15.209, 15.249, FCCID : T68-FWCS IC:6627A-FWCS
Ingress Protection (IP) rating IP54

Fluke CNX t3000 Wireless K-Type Temperature Module


Specifications
Range K-Type: -200 °C to 1372 °C
Resolution 0.1 °C
Accuracy ±[0.5 % + 0.3 °C] counts
Input terminals K-Type mini
LCD w/backlight 3 digits
Log rate/interval 1 sec minimum/adjustable by PC
Battery type 2 AA, NEDA 15 A, IEC LR6
Battery life 400 hours
Memory Record up to 65,000 readings
RF communications 2.4 GHZ ISM Band
RF communication range 20 Meters
Operating temperature -10 °C to +50 °C
Storage temperature -40 °C to +60 °C
Temperature scale ITS-90
Operating humidity 90 % at 35 °C, 75 % at 40 °C, 45 % at 50 °C
EMC EN 61326-1:2006
Safety compliance CAT I per IEC/EN 61010-1:2010
EN/IEC 61010-2-030:2010
Pollution degree 2
Temperature coefficient 0.01 % of reading +0.03 °C per °C
Certifications CSA, FCC T68-FWCS IC:6627A-FWCS
Ingress Protection (IP) rating IP42 with thermocouple attached 
Size (HxWxD) 16.5 cm x 6.35 cm x 3.56 cm (6.5 in x 2.5 in x 1.4 in)
Weight .22 kg (8 oz)
 

Model Name Description
CNX t3000 Kit Fluke CNX t3000 Temperature Measurement Kit
  • Fluke CNX 3000 Wireless Multimeter
  • Fluke CNX t3000 K-Type Wireless Temperature Module
  • TL175 Test Leads
  • AC175 Alligator Clips
  • 80PK-1 K-Type Bead Thermocouple
  • Magnetic Hanging Strap
  • Soft Carrying Case
CNX v3000 Kit

Fluke CNX v3000 AC Voltage Kit

  • Fluke CNX 3000 Wireless Multimeter
  • Fluke CNX v3000 AC Wireless Voltage Module
  • TL224 Test Leads
  • TL175 Test Leads
  • AC175 Alligator Clips
  • AC285 Alligator Clips
  • Magnetic Hanging Strap
  • Soft Carrying Case
CNX a3000 Kit Fluke CNX a3000 AC Current Clamp Kit
  • Fluke CNX 3000 Wireless Multimeter
  • Fluke CNX a3000 AC Wireless Current Clamp Module
  • TL175 Test Leads
  • AC175 Alligator Clips
  • Soft Carrying Case
CNX i3000 Kit Fluke CNX i3000 iFlex AC Current Measurement Kit
  • Fluke CNX 3000 Wireless Multimeter
  • Fluke CNX i3000 iFlex AC Wireless Current Module
  • iFlex i2500-10 Flexible Current Probe
  • TL175 Test Leads
  • AC175 Alligator Clips
  • Magnetic Hanging Strap
  • Soft Carrying Case
CNX 3000 Fluke CNX 3000 Wireless Multimeter
  • True-RMS Wireless Multimeter
  • TL175 Test Leads
  • AC175 Alligator Clips
CNX v3000 Fluke CNX v3000 AC Wireless Voltage Module
  • True-RMS AC Voltage Module
  • TL224 Test Leads
  • AC285 Alligator Clips
  • Magnetic Hanging Strap
CNX i3000 Fluke CNX i3000 iFlex AC Wireless Current Module
  • True-RMS Flexible Current Meter
  • iFlex i2500-10 Flexible Current Probe
CNX a3000 Fluke CNX a3000 AC Wireless Current Clamp Module
  • True-RMS AC Clamp Meter
CNX t3000 Fluke CNX t3000 K-Type Wireless Temperature Module
  • K-Type Thermocouple Thermometer
  • 80PK-1 K-Type Bead Thermocouple
  • Magnetic Hanging Strap
CNX 3000 GM Fluke CNX 3000 General Maintenance System
  • Fluke CNX 3000 Wireless Multimeter
  • Fluke CNX v3000 AC Wireless Voltage Module
  • Fluke CNX i3000 iFlex AC Wireless Current Module
  • iFlex i2500-10 Flexible Current Probe
  • Fluke CNX pc3000 Wireless PC Adapter and Software
  • TL224 Test Leads
  • TL175 Test Leads
  • AC175 Alligator Clips
  • AC285 Alligator Clips
  • Two (2) Magnetic Hanging Straps
  • Soft Carrying Case
CNX 3000 IND Fluke CNX 3000 Industrial System
  • Fluke CNX 3000 Wireless Multimeter
  • Fluke CNX v3000 AC Wireless Voltage Module
  • Three (3) Fluke CNX i3000 iFlex AC Wireless Current Modules
  • Three (3) iFlex i2500-10 Flexible Current Probes
  • Fluke CNX pc3000 Wireless PC Adapter and Software
  • TL224 Test Leads
  • TL175 Test Leads
  • AC175 Alligator Clips
  • AC285 Alligator Clips
  • Four (4) Magnetic Hanging Straps
CNX 3000 HVAC Fluke CNX 3000 HVAC System
  • Fluke CNX 3000 Wireless Multimeter
  • Fluke CNX a3000 AC Wireless Current Clamp Module
  • Fluke CNX t3000 K-Type Wireless Temperature Module
  • Fluke CNX pc3000 Wireless PC Adapter and Software
  • TL175 Test Leads
  • AC175 Alligator Clips
  • 80PK-1 K-Type Bead Thermocouple
  • Magnetic Hanging Strap
  • Soft Carrying Case

Cases and Holsters
CNX C3000 CNX C3000 Premium Modular Tool Bag
Webbed bag for customizable carrying system, when used with other CNX modular cases.
CNX C3001 CNX small modular Case C3001
Compatible with wireless modules Fluke CNX
CNX C3002 CNX C3002 Modular DMM 2-Compartment Soft Case
  • Compatible with Fluke CNX Wireless Multimeter
  • Inner pocket and carrying handle
CNX C3003 CNX C3003 Modular 3-Compartment Soft Case
  • Compatible with Fluke CNX wireless modules and meters
  • Three adjustable compartments, inner pocket, and carrying handle

PC Adapter and Software
CNX pc3000 Fluke CNX pc3000 PC Adapter
CNX PC Adapter Software

Temperature
80PJ-1 80PJ-1 Bead Probe
80PK-1 80PK-1 Bead Probe

Test Leads
TL175 TL175 TwistGuard™ Test Leads
 

 

More detail about Pacific Radio (M) Sdn Bhd
Pacific Radio (M) Sdn Bhd
Pacific Radio (M) Sdn Bhd Measuring Instruments Supplier Malaysia, Soldering Irons and Switches Supplier Kuala Lumpur, KL, Audio Equipments Supplier Singapore ~ Pacific Radio (M) Sdn Bhd
Contact Us flagMalaysia