Technical Data
Differential Pressure - Piezoresistive |
Measuring range |
0 to 300 hPa |
Accuracy |
±0.5 hPa (0.0 to 50.0 hPa), ±1 % of mv (50.1 to 100.0 hPa), ±1.5 % of mv (Remaining Range) |
Resolution |
0.1 hPa |
Flue gas O₂ |
Measuring range |
0 to 21 Vol.% |
Accuracy |
±0.2 Vol.% |
Resolution |
0.1 Vol.% |
Flue gas CO (with H₂-compensation) |
Measuring range |
0 to 8000 ppm |
Accuracy |
±10 ppm or ±10 % of mv (0 to 200 ppm), ±20 ppm or ±5 % of mv (201 to 2000 ppm), ±10 % of mv (2001 to 8000 ppm) |
Resolution |
1 ppm |
CO determination (with H₂-compensation), automatic dilution |
Measuring range |
0 to 30000 ppm |
Accuracy |
±200 ppm or ±20 % of mv (0 to 30000 ppm) |
Resolution |
1 ppm |
Flue gas NO |
Measuring range |
0 to 3000 ppm |
Accuracy |
±5 ppm (0 to 100 ppm), ±5 % of mv (101 to 2000 ppm), ±10 % of mv (2001 to 3000 ppm) |
Resolution |
1 ppm |
Option
Flue gas Draught |
Measuring range |
-9.99 to +40 hPa |
Accuracy |
±0.02 hPa or ±5 % of mv (-0.50 to +0.60 hPa), ±0.03 hPa (+0.61 to +3.00 hPa), ±1.5 % of mv (+3.01 to +40.00 hPa) (the greater value applies) |
Resolution |
0.01 hPa |
Temperature |
Measuring range |
-40 to +1200 °C |
Accuracy |
±0.5 °C (0 to +100.0 °C), ±0.5 % of mv (Remaining Range) |
Resolution |
0.1 °C(-40 to +999.9 °C), 1 °C (Remaining Range) |
dependent on the thermocouple in the flue gas probe
Flue gas degree of effectivity, Eta (calculated) |
Measuring range |
0 to 120 % |
Resolution |
0.1 % |
Flue gas loss (calculated) |
Measuring range |
0 to 99.0 % |
Resolution |
0.1 % |
Flue gas CO₂ calculation (calculated from O₂) |
Measuring range |
Display range 0 to CO₂ max |
Accuracy |
±0.2 Vol.% |
Resolution |
0.1 Vol.% |
General technical data |
Dimensions |
270 x 160 x 57 mm |
Operating temperature |
-5 to +45 °C |
Power supply |
Rech. batt. block 3.7 V / 2.6 Ah; Mains unit 6 V / 1.2 A (optional) |
System requirements |
requires iOS 7.1 or newer; requires Android 4.3 or newer; requires mobile end device with Bluetooth 4.0 |
Maximum memory |
500,000 readings |
Storage temperature |
-20 to +50 °C |
Weight |
720 g (excluding battery) |
Applications
Differential pressure measurement
You can use the testo 330i flue gas analyzer to check the gas supply on gas boilers. You can therefore measure gas flow pressure and gas resting pressure in order to ensure that the system can be commissioned. You can also use the data specified by the manufacturer for the nozzle pressure to tailor the appliance output to the heat requirement.
Flue gas analysis and flue gas measurement
The testo 330i is equipped with two long-life sensors for measuring O2 and CO and can optionally be upgraded with an NO sensor. Flue gas and draught are measured in parallel, so all relevant flue gas parameters are displayed simultaneously on the smartphone.
O2 supply air measurement
In balanced flue systems, you can use the testo 330i flue gas analyzer to check whether there is a leak in the flue gas system. Measure the O2 content of the combustion air in the dual wall clearance to ensure that no flue gas is getting into the combustion air supply, thus preventing damage due to condensate or impairment of the combustion process.