Item Code | 211210 |
---|---|
Model | INC820 |
System (Upper Chamber) | Natural convection, Air jacket |
Operating Temp. Range(Upper Chamber) | Room temp.+5℃~80℃ |
Temp. Adjustment Accuracy (Upper Chamber) | ±0.5℃(at 37℃) |
Temp. Distribution Accuracy (Upper Chamber) | ±1℃(at 37℃) |
Interior | Stainless Steel |
Heater (Upper Chamber) | Ferrochrome Heater |
Heat Insulation Material (Upper Chamber) | Glass Wool |
Heater Power (Upper Chamber) | 400W |
Cable Port (Upper Chamber) | 30mm ID (Right side from the front) |
Exhaust Port (Upper Chamber) | 30mm(with Damper), 2 ports on the top |
Controller (Upper Chamber) | VS3 Type, Fixed operation temperature controller |
Operation Function (Upper Chamber) | Fixed temp., Quick auto stop, Auto stop, Auto start operation |
Internal Dimensions (Upper Chamber) | 600×530×500mm |
Internal Capacity (Upper Chamber) | 150L |
Shelf Step/ Pitch (Upper Chamber) | 13 steps / 30mm |
Standard Shelf / Support (Upper Chamber) | 2 shelves/4 suppots |
System (Lower Chmber) | Forced convection |
Operating Temp. Range(Lower Chamber) | 4℃~50℃ |
Temp. Adjustment Accuracy (Lower Chamber) | ±0.3℃(at continuous refrigeration) |
Temp. Distribution Accuracy (Lower Chamber) | ±1℃(at continuous refrigeration) |
Interior | Stainless Steel |
Heater (Lower Chamber) | Ferrochrome Heater |
Heat Insulation Material (Lower Chamber) | Stylene foam (Non-freon) |
Heater Power (Lower Chamber) | 550W |
Blower (Lower Chamber) | Axial fan |
Refrigerator (Lower Chamber) | Air cooling, 250W |
Refrigerant (Lower Chamber) | R404A |
Defrost Operation (Lower Chamber) | Manual / Cycle |
Cable Port (Lower Chamber) | 30mm ID (Right side from the front) |
Controller (Lower Chamber) | VS4 Type Program operation temp. controller |
Operation Function (Lower Chamber) | Program operation(30 steps ×1、15 steps×2、10 steps×3), Fixed temp. operation, Quick auto stop, Auto stop, Auto start |
Safety Device (Lower Chamber) | Self-diagnostic functions(Temp. sensor error, Heater disconnection, SSR short-circuit, Main relay error, Automatic overheat prevention), Key-lock, Over current ELB, Overheat preventive device |
Interior (Lower Chamber) | 600×477×500mm |
Internal Capacity | 143L |
Shelf support step number/pitch | 13 steps / 30mm |
Shelf / Support | 3 shelves/6 pcs. |
Sensor | Pt100Ω |
Option | RS485 Terminal, Recorder contact terminal, Alarm terminal |
Shelf Loading Capacity | Approx. 15kg/ shelf |
Door Key | 1 pc. each |
External dimensions(W×D×H mm) | 710×656×1792mm |
Power Source |
AC115/220V Single phase |
Weight | 160kg |