Extech PQ34-2 200A Current Clamp Probes

Extech PQ34-2 200A Current Clamp Probes

分类: Power Analyzers 已售罄
欲了解更多详情,请游览我们的官方网站 lelab.com.my

详情

Specifications Range
Measurement ranges 20A and 200A, switch selectable
Output sensitivity 10mV/A (20A range)
1mV/A (200A range)
Output coupling AC coupled.
Accuracy (50/60Hz) ± 1% F.S. @ 23 ± 5°C (50/60Hz)
Frequency range 45 to 400 Hz.
Phase error (50/60Hz) <± 1° (50/60Hz)
Minimum load 100 K ohm for specified accuracy
Enclosure material UL 94-V0 rated
Safety IEC 1010, Category III 600V
Power Supply 8V± 1V (supplied from meter)
Max. Jaw Open Size 19 mm (0.75 inch) Dia.
Operating Temperature 0°C to 50°C (32°F to 122°F)
Operating Humidity Less than 80% RH
Dimension 179.9 x 66.9 x 40.1mm. (7.0 X 2.6 X 1.57''inch)
Cable length : 4 meters (13.1 feet)
DC plug size : 5.5 mm x 2.5 mm DIA.
Weight 330 g (0.73lb)

更多 LELab Sdn Bhd 相关资料