UV Accelerated Weather Aging Test Chamber ZY-D

UV Accelerated Weather Aging Test Chamber ZY-D

分类: Environmental & Weather Aging Testing 当前有货
可在官方网站购买
欲了解更多详情,请游览我们的官方网站 carlssoon-tech.com.my

详情

UV Accelerated Weather Aging Test Chamber
Brand: Rankconn Instruments

Model: ZY-D

 
 
Product Introduction: 
UV accelerated weathering aging test chamber can provide reliable aging test data to make the exact prediction of products' weathering fastness (aging resistance), which is conducive to sift and optimize the formula. UV accelerated weathering aging test chamber is used in many industries, such as: paint, inks, resin, plastics, printing and packaging, adhesives, auto and motorcycle industry, cosmetics, metal, electronic, electroplating, medicine, etc.
 
 
 
Usage:
UV accelerated weathering aging test chamber is the most widely used UV Weathering tester in the world, there are thousands of ultraviolet aging testers coming into use. UV accelerated weathering aging test chamber has become the standard of the UV test, fit standards of ISO, ASTM, DIN, JIS, SAE, BS, ANSI, GM, U.S.GOVT, etc. It can simulate the damage of sunshine, rain, dew to the material quickly and exactly. Damage types including bleaching, discoloration, loss of light, powder, crack, turbidity, bubble-based, embrittlement, strength decline and oxidation need to be engendered in months or years but can engender in a few days or weeks.
 
 
Standards:
IEC61215, ASTM D4329, D499, D4587, D5208, G154, G53;
ISO 4892-3, ISO 11507; EN 534; BS 2782; JIS D0205; SAE J2020, etc.
 
 
Features:

 
 
Technical parameters:
 

Chamber dimension 450*1170*500mm
Materials Inside and Outside SUS#304 stainless steel
Temperature range RT+10~70°C
Temperature uniformity ±2°C
Temperature fluctuation ±0.5°C
Temperature control PID SSR control
Humidity range: 95%RH
Controller Korean TEMI 880 programmable controller
Control mode Balance temperature humidity control (BTHC)
Test cycle setting Illumination, condensation and water spray test cycle is programmable
Lamp power 40W/Piece
Distance from sample to lamp 50±2mm (adjustable)
Centre distance between the lamp 70mm
Wavelength ultraviolet light UV-A: 315-400nm; UV-B: 280-315nm for optional (8pcs, 2000h lifetime)
Irradiation black panel temperature 45°C~80°C
Standard Specimen Size 48* 75×150mm (special specifications can be custom-made)
Water consumption 8L per day
Testing time 0~999H, adjustable
Power 220V/50Hz /±10% 4.5KW
Protection system Overload short circuit protection, over temperature protection, water lacking
 
 
 
 
 
 
 