AirFlow HVS-TSP High Volume Sampler

AirFlow HVS-TSP High Volume Sampler

分类: AMS Analitica 当前有货
欲了解更多详情,请游览我们的官方网站 sciencegates.com

详情

AirFlow HVS-TSP High Volume Sampler
Sampler for environmental parameters

  • Flow range 120-1400 liters/minute
  • Venturi electronic flow control system (ISO-5167)
  • High resolution large display LCD backlight RS 232 PC interface and USB port for easy files download
  • Sampling period are microprocessor based with standard deviation control and pressure drop check on filter
  • Wind velocity and direction connection interface included (no sensors)
  • Remote data handling with GSM/GPRS protocols (Optional)
  • External shelder in anodized aluminium for outdoor applications
  • Ambient temperature and barometric pressure detected
  • Adapter kit available to convert Puf sampler in high volume PM sampler
  • Class protection IP55
  • Adapter low ambient temperature kit (Max.-30°C)


Sampling Features

The Air Flow TSP high volume sampler with the new PC electronic board allow the user to view all setting and sampling data for a full compliance sampling status. The follolwing data are available at display in a different screens:

- Selected sampling time (Max 9999 minute)
- Residual sampling time to stop
- Ambient Temperature
- Sampler cabinet temperature
- Barometric atmospheric pressure
- Sampling flow-rate
- Total sampling volume
- Total standardized sampling volume
- Sampling faults (flow or sensors readings)
- Wind velocity and direction (with optional sensors installed)
- Stored sampling data

The new microprocessor capabilities allow the user to select and store in the memory up to 3 different programs with sampling settings ready to start. Four different parameters are available on the main screen in a graphic format as Ambient Temperature, Barometric Pressure, Flowrate and Standard Deviation. Automatic re-start after power shut-off with all data stored in the memory.
 

Other Technical Features and Part numbers

-Flow range 120-1400 lit/min
- Pump type: Three stage brushless motor
- Maintenance period: Up 20.000 hours depending on work conditions
- Power source: 220/240 V-AC 50 Hz
- Power Comps. (Ampere) 15 at start / 4 during sampling procedures
- Operative conditions -5° / +50° (-30°C with heating conditioning system)
- Dimensions 120x37x37 cm (hxlxp)
- Weight (Kg.) 14 Kg


Order Information:

Code Product
600/AFTPS1001K Air Flow TSP high volume sampler
600/GRMET001K Wind speed and direction sensor with assembly kit
600/AFPUF1001GSM GSM modem
600/SCS001K Heating conditioning system (Max.-30°C)
NFS/GF1-8X10-100 Glass Fiber Filter Grade GF1 (Eq.GFA) 20x25 (8x10 inch) box of 100
FAN/MFQ203254 Quartz Filter 8x10 box of 25pcs

更多 SCIENCEGATES SDN BHD 相关资料
SCIENCEGATES SDN BHD
SCIENCEGATES SDN BHD Laboratory Equipment Supplier Selangor, Scientific Instruments KL, Lab Water Purification Systems Malaysia ~ SCIENCEGATES SDN BHD