HIOKI RESISTANCE METER RM3545A

HIOKI RESISTANCE METER RM3545A

分类: Resistance Meter 当前有货
可在官方网站购买
欲了解更多详情,请游览我们的官方网站 accutac.com.my

详情

HIOKI RESISTANCE METER RM3545A

New Heights in 100% Inspection.
Market leading precision tests for testing every weld or connection on your production line.
 

As electric mobility gains popularity, manufacturers are developing components for higher currents and voltages. Since even minuscule amounts of resistance can have a significant impact on energy efficiency and safety, more accurate quality control focusing on resistance is required. The Resistance Meter RM3545A facilitates accurate resistance measurements and is suitable for diverse applications in development and production lines.

 

Manage connection quality in welded materials and other parts quantitatively
Quantitatively verify weld quality and weld methods in EV power cables and other parts
Use readings as indicators for thermal design and energy management
Use accurate resistance measurements to simulate heat loss and energy efficiency
Boost productivity by embedding the instrument in automatic test equipment
Embed the instrument in a system without needing to worry about wiring resistance or contact resistance. The instrument is ideal for use in high-speed 100% inspections

Key Features

 

Basic specifications

Accuracy guaranteed: 1 year
Resistance range (13 ranges) [Range, Maximum display value, Resolutions, Testing current (Measurement current) ]
1000 μΩ: 1200.000 μΩ, 1 nΩ, 1 A
10 mΩ: 12.000 00 mΩ, 10 nΩ, 1 A
100 mΩ: 120.000 0 mΩ, 100 nΩ, 1 A
1000 mΩ: 1200.000 mΩ, 1 μΩ, 100 mA
10 Ω: 12.000 00 Ω, 10 μΩ, 10 mA
100 Ω: 120.000 0 Ω, 100 μΩ, 10 mA
1000 Ω: 1200.000 Ω, 1 mΩ, 1 mA
10 kΩ: 12.000 00 kΩ, 10 mΩ, 1 mA
100 kΩ: 120.000 0 kΩ, 100 mΩ, 100 μA
1000 kΩ: 1200.000 kΩ, 1 Ω, 10 μA
10 MΩ 12.000 00 MΩ, 10 Ω, 1 μA
100 MΩ (100 MΩ range high-precision mode): 120.000 0 MΩ, 100 Ω, 100 nA
1000 MΩ: 1200.0 MΩ, 100 kΩ, 1 μA or less
Representative accuracy (High mode, OVC function enabled, SLOW2, no zero adjustment) 1000 μΩ range: ±0.045% rdg ±0.010 % f.s.
10 mΩ range: ±0.045% rdg ±0.001 % f.s.
100 mΩ range: ±0.045% rdg ±0.001 % f.s.
1000 mΩ range: ±0.012% rdg ±0.001 % f.s.
1000 Ω range: ±0.006% rdg ±0.001 % f.s.
Testing current (Measurement current) High mode: 1000μΩ (1 A) to 1000 MΩ (up to 1 μA)
Low mode: 100 mΩ (100 mA) to 100 Ω (1 mA)
Low power mode (LP): Low power measurement with measurement current and open circuit voltage down to 20 mV
LP1000 mΩ (1 mA) to LP1000 Ω (5 μA)
Measurement speed Representative value: FAST (2.3 ms) / MED (50 Hz: 22 ms, 60 Hz: 19 ms) / SLOW1 (102 ms) / SLOW2 (202 ms)
PR10 mΩ range*: FAST (21 ms) / MED (50 Hz: 41 ms, 60 Hz: 37 ms) / SLOW1 (121 ms) / SLOW2 (221 ms)
Path resistance tolerance (reference values)
Path resistance between SOURCE B and SOURCE A (other than measurement target)
Range: 100 mΩ or less (PR* mode off) : 2.6 Ω
Range: 100 mΩ or less (PR* mode on) : 3.5 Ω
Range: 1000 mΩ: 15 Ω
Range: 10 Ω : 150 Ω
Range: 100 Ω : 100 Ω
Range: 10 kΩ : 500 Ω
Range: 100 kΩ or greater: 1 kΩ
Maximum open-terminal voltage Range: 1000 Ω or less: 8.0 V
Range: 10 kΩ or greater: 20 V
Temperature measurement Temperature Sensor (Z2001[included accessories]): -10.0 to 99.9
Analog input (Ex: Infrared thermometer): DC 0 V to 2.0 V
Multiplexer (built-in option)*
*RM3545A-2 only
Multiplexer unit Z3003
Number of installable unitsMax. 2
Maximum number of channels : 20 channels (4-wire method), 42 channels (2-wire method)
Switching time : 30 ms
Multiplexer (external option) Switch Mainframe
Maximum number of channels (SW1001): 33 channels (4-wire method)
Maximum number of channels (SW1002): 132 channels (4-wire method)
Switching time11 ms
Communication interfaces LAN (TCP/IP, 10BASE-T/100BASE-TX), RS-232C (Max. 115200 bps, also used as printer interface), USB, EXT I/O (D-sub 37-pin, Analog output (D/A output voltage range: DC 0 V to 1.5 V )
Functions Contact check, Zero adjustment (within each range ±50% f.s.) *1, Zero-adjustment-free accuracy guaranteed, OVC function, Contact improvement function (max. applied voltage: 5V; max. applied current: 10 mA), Low-power mode (maximum open voltage: 20 mV), Auto-hold function, Comparator, Temperature measurement function, Temperature correction (TC) function, Temperature conversion (ΔT) function, Statistical calculation function, Delay function, Averaging function, Saving panels (saving of measurement conditions), Data memory function, Command monitor function (display of send/receive status of commands and queries),LabVIEW® Driver compatible*2
*1: Zero adjustment forcibly disabled for 100 MΩ or greater
*2: LabVIEW Driver is the trademark or registered trademark of National Instruments.
Power supply 100 V to 240 V AC, 50 Hz/60 Hz, Rated power consumption: 40 VA max.
Normal power consumption (Reference value) 16 W (Testting current 1 A, LCD on)
Dimensions and mass RM3545A-1: 215 mm (8.46 in) W × 80 mm (3.15 in) H × 306.5 mm (12.07 in) D, 2.7 kg (95.2 oz)
RM3545A-2: 215 mm (8.46 in) W × 80 mm (3.15 in) H × 306.5 mm (12.07 in) D, 3.4 kg (119.9 oz)
Included accessories Power cord ×1, Temperature sensor Z2001 ×1, Male EXT I/O connector ×1, EXT. I/O connector cover ×1, Spare fuse (F1.6AH 250 V)×1, Start up guide ×1, Operating Precautions ×1, Instruction manual ×1

*PR: Pure Resistance

更多 Accutac Sdn Bhd 相关资料
Accutac Sdn Bhd
Accutac Sdn Bhd Thermal Imager Supplier Penang, Data Logger, LCR Meter Supply Malaysia, Power Quality Meter, Precision Measuring Instruments Supplies Bayan Lepas ~ Accutac Sdn Bhd