TUNKIA - TD1230 Verification Device for Withstanding Voltage Tester

TUNKIA - TD1230 Verification Device for Withstanding Voltage Tester

分类: 当前有货
欲了解更多详情,请游览我们的官方网站 carlssoon-tech.com

详情

TD1230 Verification Device for Withstanding Voltage Tester
Brand: Tunkia
Model: TD1230

 

Product Definition:
TD1230 applied to calibrate AC/DC withstanding voltage tester, withstanding voltage part of safety performance comprehensive tester, withstanding voltage part of insulation withstanding voltage tester and etc.
 
1. Summary
TD1230 applied to calibrate AC/DC withstanding voltage tester, withstanding voltage part of safety performance comprehensive tester, withstanding voltage part of insulation withstanding voltage tester and etc.
 
2. Features
• AC/DC voltage measurement100 V15 kV.
• AC/DC current measurement0.1 mA250 mA.
• Distortion / Ripple function and voltage duration time test function.
• Built-in adjustable resistive load with fineness of 1 kΩ.
• Optional high power load box (TD1240) help calibrate the output capacity.
• Oscilloscope and data statistical analysis function.
• LCD touch screen.
• Setting output values by keys.
• Notification of wire connection.

3. Applications


5. Specifications:

5.1 AC/DC Voltage Measurement

Range Resolution Accuracy
±(ppm of reading + ppm of range) [1]
Class 0.5 Class 0.2
5 kV 0.1 V 4000 + 1000 1800 + 200
10 kV[2] 1 V 4000 + 1000 1800 + 200
15 kV[2] 1 V 4000 + 1000 1800 + 200
Note [1] : (ppm = parts per million) (e.g.10ppm = 0.001%).
Note [2]: 10 kV, 15 kV ranges are selectable
• Measuring range: 100 V to 15 kV, 5-bit display, manual/automatic range switching
• Input impedance: 50 MΩ, true RMS response for AC, average response for DC

5.2 AC/DC Current Measurement

Range Resolution Accuracy
±(ppm of reading + ppm of range) [1]
Class 0.5 Class 0.2
2 mA 0.1 μA 4000 + 1000 1800 + 200
20 mA 1 μA 4000 + 1000 1800 + 200
200 mA 10 μA 4000 + 1000 1800 + 200
• Measurement range: 0.1 mA to 250 mA, 5-digit display, manual/automatic range switching
• Built-in adjustable resistive load of 6 kΩ8MΩ with fineness of 1 kΩ
• True RMS response for AC, average response for DC

5.3 Other Functions

Parameters Range Resolution Accuracy
AC Frequency 45 Hz65 Hz 0.1 Hz ± 0.2%
AC Voltage Distortion 0.5%10% 0.01% ± 0.5%
DC Voltage Ripple 0.1%30% 0.01% ± 0.2%
Voltage Duration Time 0.1 s1000 s 0.001 s AC± (0.2%*RD + 2 periods)
DC± (0.2%*RD + 40 ms)


6. General Specifications
Power Supply AC ( 220 ± 22 ) V( 50 ± 2 ) Hz
Temperature Performance Working temperature0~45;
Storage temperature-20~70
Humidity Performance Working humidity< 80% @ 30< 70% @ 40< 40% @ 50
Storage humidity(20%~80%) R·Hnon-condensing
Interface RS232