Designed with user functionality and performance in mind, the HI935007 is a waterproof K-Type thermocouple thermometer that measures temperature from -50.0 to 300°C (-58.0 to 572°F). The included batteries provide up to 3500 hours of continuous use. CAL Check™ and the Battery Error Prevention System (BEPS) ensure that the readings are accurate. The probe is permanently attached to the meter and is not replaceable
The HI935007 is a waterproof portable K-Type thermocouple thermometer made for the food professional that is required to monitor temperature as part of a hazardous analysis of critical control points (HACCP) plan including in food service, production, packaging, transportation, restaurants or catering. The HI935007, as a meter, can measure over a wide range of temperatures from -50.0 °C (-58.0 °F) up to 300 °C (572 °F) and offers a very high accuracy of ±0.4 °C (±0.7 °F). The accuracy of the meter is assured with advanced diagnostic features including CAL-Check that checks for abnormal drift of the internal electronics, battery error prevention system (BEPS) that will not allow a reading to be taken when there is not sufficient battery power and probe diagnostics that alert the user when the probe is not connected or has been damaged.
HI935007 Foodcare thermometer is supplied with a fixed attached penetration K-Type thermocouple probe that will reach 90% of the final reading within 15 seconds. The tip of FC766PW is just 3 mm (0.12'') in diameter allowing for easy penetrations into solids and semi-solids. The AISI 316 stainless steel body is 155 mm (6.1'') long and is safe for food contact in compliance with Regulation (EC) 1935/2004.
There are many considerations when designing a thermometer for the food professional. Below are features that we considered to be important and beneficial.
The calibration check (CAL Check) feature of the HI935007 is an internal diagnostic feature that checks for any drift in the electronics that occurs with all digital thermometers over time. When the meter is turned CAL Check looks to see if the internal calibration is within +/- 0.3 °C. If the drift is greater and error (err) message will be displayed. With CAL Check you can be confident that the meter is working properly.
The included AAA batteries provide 3500 hours of continuous use. When the battery level is below 10%, a warning symbol will blink to indicate low battery condition. When low battery power may affect your results, the meter will shut down to avoid erroneous readings caused by low battery level.
If the probe is damaged the message will be displayed along with four blinking dashed lines. For an accurate temperature measurement it is important to know that not only is the meter working properly but that the probe is okay.
The attached penetration K-Type thermocouple probe that will reach 90% of the final reading within 15 seconds.
The tip of the probe is just 3 mm (0.12'') in diameter allowing for easy penetrations into solids and semi-solids.
The AISI 316 stainless steel body is 155 mm (6.1'') long and is safe for food contact in compliance with Regulation (EC) 1935/2004.
Supplied with Instrument Quality Certificate
Large LCD
Stability Indicator
Automatic Shut-off
Ergonomic, rugged, waterproof (IP65) design
SKU | HI935007 |
---|---|
Product Name | Foodcare K-Type Thermocouple Thermometer with Fixed Attached Probe - HI935007 |
Quote Required | Yes |
K-Type Thermocouple Temperature Range |
-50.0 to 199.9 oC; 200 to 300 oC -58.0 to 399.9 oF; 400 to 573 oF |
K-Type Thermocouple Temperature Resolution |
0.1 oC (-50.0 to 199.9 oC); 1 oC (200 to 300 oC) 0.1 oF (-58.0 to 399.9 oF); 1 oF (400 to 573 oF) |
K-Type Thermocouple Temperature Accuracy |
+/- 0.4 oC (-50.0 to 300 oC) +/- 0.7oF (-58.0 to 573 oF) |
Temperature Probe | K-Type attached penetration probe (fixed) |
Automatic Shut-Off | user selectable |
Battery Type/Life | 1.5V AAA battery (3 pcs.) / approximately 3500 hours of continuous use |
Environment | operating conditions |
Dimensions | 140 x 57 x 28 mm (5.5 x 2.2 x 1.1”) |
Weight | 178 g (6.27 oz.) |
Ordering Information | HI935007 is supplied with K-Type thermocouple penetration probe, batteries, instruction manual, and instrument quality certificate. |