The XGS-600 Gauge Controller accepts any of three different kinds of gauge cards which, between them, operate all Agilent vacuum gauges. Choose from
Up to six cards can be installed in the XGS-600 Controller, in practically any combination to suit your particular application. Everything else is standard equipment. This yields maximum flexibility for matching the system to your vacuum measurement requirements. The XGS-600 Gauge Controller allows for simultaneous operation of up to 12 gauges and can display readings for up to 8 gauges at any one time. All settings are stored in flash memory, so no worry about reprogramming after a power outage.
User interface | LCD monochrome dot matrix, 240 x 128, white LED backlight; 5 button keypad |
Display capability | Up to eight simultaneous channels of pressure measurement, up to a 5 character user label for each; up to two channels in a large font; readable from 15 feet; display updates every 200 msec |
Gauge capability | Up to 4 Hot Filament BA gauges including metal, glass, and nude UHV types using the HFIG** cards; up to 5 IMGs* using the IMG* cards; up to 12 rough gauges using the Dual Convection Gauge Cards; up to 12Active Gauges using Dual Analog Input Cards |
Measurement range | 1 x 10-11Torr to Atm dependent upon card and gauge type selected |
Setpoints | Eight normally open setpoints (open collector), with independently programmable ON and OFF delays, manual override (see detail specs for ratings) |
Fast setpoint response | 20 msec max (with delay set to 0.0 secs) |
Serial communications | RS232 and RS485 (fully programmable), Profibus DP (optional) |
Analog outputs | 1v/dec LOG, 0 – 10 V FS, on each card with 200 msec update rate |
I/O control | HFIG** card: Remote gauge ON/OFF, Fil select, DeGAS ON/OFF Inputs, Status Output IMG* card: Remote gauge ON/OFF Input, Status Output Dual Convection Card: Remote CAL inputs, Status Outputs Dual Analog Input Card: Remote CAL Inputs, Status Output |
Other | Auto-ON capability for installed ion gauges when at least one convection card is installed Sensitivity, gas correction, emission current (BA gauges only) inputs for ion gauges E-BEAM degas, with auto turn off for BA gauges Selectable air or argon calibration for rough gauges Adjustable atmosphere value for rough gauges Text error messages All data stored in FLASH non-volatile memory requiring no batteries |
Power input/Temperature | Worldwide universal power input, 100-240 VAC, 2 A, 50/60 Hz, IEC inlet; 5-40 ºC ambient |
Regulatory approvals | CE certified and RoHS compliant, BV certified for USA and Canada |