The standard IEC 60364-7-710 requires an acoustic and visual alarm system to be installed at a suitable place so that the medical staff can permanently monitor it. In most cases, the remote signalling panel is installed on the operating theatre/room panel or at the nurse station.
 

Technical Features:

Model HRI-RSCP
Facial Material Polycarbonate
Displays Digital - 7-segment LED, 3 digits
Analogue - LED bar graph, 3 colours
Parameters Display Digital - Insulation Resistance, Load Current,
Analogue - Insulation Resistance, Load Current
Alarm Signalling Buzzer and LEDs
Buttons TEST and MUTE (also to toggle between load current and transformer temperature)
Serial Interface RS 485 Modbus RTU
Power Supply DC 12 V
Dimensions (W x D x H) / Weight 125mm x 40mm x 80mm / 200g
Standards/Certifications MS IEC/IEC 60364-7-710, EN 55011, EN 61000-4-2, EN 61000-4-3, EN 61000-4-8, EN 61000-4-11, CE, TUVNORD, SIRIM