Speed, Sensitivity and Reliability for Demanding Leak Detection Applications
The INFICON UL1000 Mobile Helium Leak Detector is designed to meet the most demanding industrial leak detection applications. Providing fast, accurate and repeatable test results, testing flexibility, and high sensitivity in a low maintenance design, the UL1000 is ready to tackle the toughest leak detection jobs industry can offer. The UL1000 is optimized to provide quick, accurate results in any application from the leak checking of large vessels and systems to high-cycle, repetitive component testing - 24 hours / 7 days a week - in the toughest industrial environments.
With the addition of the TC1000 Test Chamber accessory, the UL1000 Helium Leak Detector provides easy, fast and accurate testing of hermetically sealed parts like IC packages, quartz crystals and laser diodes
(according to MIL-STD 883, Method 1014).
Leak testing and quality control of all types of components, including:
Min. detectable leak rate in vacuum mode (He) | mbar•L/s | <5E-12 |
MIn. detectable leak rate in sniffing mode (He) | mbar•L/s | 5.00E-08 |
Max. Inlet pressure GROSS mode | mbar | 15 |
Max. Inlet pressure FINE mode | mbar | 2 |
Max. Inlet pressure ULTRA mode | mbar | 0.4 |
Pumping speed during evacuation | m³/h | 16 at 50 Hz |
Helium pumping speed GROSS mode | L/s | max.8 |
Helium pumping speed FINE mode | L/s | 7 |
Helium pumping speed ULTRA mode | L/s | 2.5 |
Detectable masses | 2,3,4 amu | |
Mass spectrometer Sector field | 180° | |
Filaments ion soure Iridium/Yttria coated | 2 | |
Calibrated built in test leak | mbar•L/s | E-7 |
Test port | ||
Adjustable triggers | 2 | |
Interface | RS232 | |
Chart recorder output | V | 2x10 |
In/Outputs | PLC compatible | |
Permissable ambient temperature (during operation ) | °C | +10.....+40 |
Weight | kg lb |
110 242 |
Dimensions ( L x W xH ) | mm in |
1068x525x850 42x21x33 |
Supply voltages | V (ac) V (ac) V (ac) |
230(+-10%)50Hz 115(+-10%)60Hz 100V(+-10%)50/60Hz |
Power consumption | VA | 1100 |