F701-S is a weighing indicator with high specifications and various functions such as super-low-drift high-precision amplifier, fast-sampling, weighing sequence functions that are useful for automatic weighing machines and control I/O.
AnalogSettingExternal signalInterfaceDisplayGeneral specificationAttachment
Analog | |
Excitation voltage | DC5V±5% Output currentwithin 90mA Ratio metric type (Up to six 350Ω load cells can be connected in parallel.) |
---|---|
Signal input range | 0.53.0mV/V , 0 to 3.0mV/V (when OIML R76-1-compliant) |
Zero adjustment range |
Automatic adjustment by digital operation 0.22.0mV/V , 0 to 2.0mV/V (when OIML R76-1-compliant) |
Span adjustment range |
Automatic adjustment by digital operation |
Minimum input sensitivity | 0.15μV/count , 0.5μV/count (when OIML R76-1-compliant) |
Accuracy |
|
A/D converter |
Conversion rate…300 times/sec. Conversion resolution…24bitbinary |
Minimum indication resolution | 1/10000, 1/6000 (when OIML R76-1-compliant) |
Secondary calibration | Equivalent input calibration Minimum indication resolution for secondary calibration1/1000ordinary temperature |
Setting | |
Setting method |
Settings are made by operating the membrane keys. Also, settings can be made from a host computer through the RS-485 interface. |
---|---|
Memory of set value |
Calibration value and a part of set value: NOV.RAM (nonvolatile RAM) Other set values: F-RAM (nonvolatile RAM) |
Protect of set value | Protect can be set by Lock switch and Lock parameter. |
Setting item |
Upper limitLower limitNear zeroSet point 1Set point 2CompensationOverUnderFinalComparison inhibit timeJudging timeComplete output timeCompensation feeding timeAuto free fall compensation regulationWeighing function 1Weighing function 2Tare settingTare displayDigital low pass filterMoving average filterMDZT periodZT rangeTotal comparison selectionTotal limit (most significant 4 digits)Total limit (least significant 5 digits)Count limitWeighing function 3Key invalid・LockInput selection 1Input selection 2Output selection 1Output selection 2Error output selectionReserve output selectionPasswordBag clamp output timeDischarging timeWeighing start timeAZ timesJudging timesSequence mode 1Sequence mode 2Filling enhancement valueExpanded function selection 1Expanded function selection 2Balance weight valueCapacityMin. scale divisionDZ regulation valueDisplay selection 1Gravitational accelerationNet overGross overDisplay selection 2RS-485 I/F settingRS-485 Communication typeRS-485 IDTransmission delay timeZero calibrationSpan calibrationEquivalent calibrationOption display |
External signal You can specify whether PNP (Source) type or NPN (Sink) type when order the F701-S. | |
Output signals16points |
SP1SP2SP3CompleteDischargeBag clamp Error selection 1,2Output selection 16Reserve 1,2 At signal ON, output transistor ON. External voltage must be prepared separately by customer. |
---|---|
Input signals 16points |
Operation permissionWeighing startStopDischarge commandManual dischargeDischarge gate openAccumulation clearInput selection 16Code 1,2,4 Contact (relay, switch etc.) or non-contact (transistor, open collector etc.) can be connected. External voltage must be prepared separately by customer. |
Interface | |
RS-485 communication interface (Select from Modbus-RTU and original format) 232RS-232C communication interfaceoption BCOBCD parallel data output interface (sink type) (option) BSCBCD parallel data output interface (source type) (option) DACD/A converter (option) ODNDeviceNet interface (option) PRFPROFIBUS interface (option) PROFIBUS-DPV0 CCLCC-Link interface (option) |
Display | |
Display |
18.5mm in character height, Numerical display on LCD(7 digit) Sub display7.3mm in character height (14 digit) |
---|---|
Indicated value | 5 digitsign: negative display at the highest digit |
Accumulation value | 9 digitThis can be changed to ''Accumulation count (4 digit)'', ''Final(5 digit)'', ''Code(1 digit)'', ''Total discharge count (6 digit)'' and ''Discharge count (5 digit)''. |
Display frequency | Select from 1, 2, 5, 10, 20 times/sec.(The system speed is 300 times/sec.) |
Capacity | 5 digit |
Min. scale division |
Setting allowable in the range of 150. * Only the following settings are valid when OIML R76-1-compliant: 0.001, 0.002, 0.005, 0.010, 0.020, 0.050, 0.01, 0.02, 0.05, 0.1, 0.2, 0.5, 1, 2, 5, 10, 20, 50 |
Decimal point | Select from 0, 0.0, 0.00, 0.000 |
Over display |
LOADA/D converter input over, OFL1Net Over, OFL2Capacity +9 scale division, -OFL2Capacity -over or -20 scale division over, OFL3Gross Over |
Center zero |
A true zero point or the center of each value is displayed. *Only true zero point is displayed when OIML R76-1-compliant. |
Status display |
CLAMPSP3SP2SP1HOLD COMPL.ZTZALMSTABTARENETGROSSNZ D.CHGHI LIMHIGOLOLO LIMCZLOCK |
General specification | |
Power supply voltage | AC100240V(1015) free power source 50/60Hz |
---|---|
Power consumption | 15W max. |
Inrush current |
2A, 3mSec:AC240V average load condition cold start at room temperature |
Operating conditions |
Temperature…Operation 10桫40桡Storage 20桫85 Humidity…85% or less (non-condensing) |
External dimensions | 192(W)×96(H)×140(D) mm (Projections excluded) |
Panel cutout dimensions |
186 +2 -0(W)×92 +1 -0(H)mm Installation panel thickness: 1.63.2mm |
Weight | Approx. 2kg |
Attachment | |
AC input cord (Nominal rating 125V) 2m | 1 |
---|---|
Loadcell connector | 1 |
Load cell connector rubber | 1 |
External I/O signal connector | 1 |
Operation manual | 1 |
Rubber packing | 1 |
BCD output connector(with BCD output option) | 1 |
DeviceNet connectorwith DeviceNet option | 1 |
CC-Link connector(with CC-Link option | 1 |
Model | Interface |
232 | RS-232C communication interface |
---|---|
BCO | BCD parallel data output interface (sink type) |
BSC | BCD parallel data output interface (source type) |
DAC | D/A converter |
ODN | DeviceNet interface |
PRF | PROFIBUS interface |
CCL | CC-Link interface |
Product | Description |
CAAC3P-P2 | AC input cord2m |
---|---|
CAAC3P-CEE7/7-P1.5 | AC input cordvoltage resistance:250V1.5m |
CA4131 | 6-wiredcable with JRC connector at one end 3m |
CA4230 | JRC-PRC 6-wiredconversion relay cable 0.3m |
CA4311 | JRC-PRC 6-wiredconversion relay cable 4-wired to 6-wiredfor 520A use1m |
CN3P-2P | 3P-2P converter plug for AC input cord |
CN10 | Loadcell connector |
CN34 | D-Sub 9p connector for RS-232 |
CN51 | BCD output connector |
CN52 | FCN series I/O connector (with cover) |
CN57 | FCN series I/O connector (with diagonal cover) |
CN71 | CC-Link connector |
CND01 | DeviceNet connector |