The HI4105 is a gas sensing, combination ion selective electrode (ISE) for the determination of carbon dioxide (CO2) in solution. Its PTFE membrane allows for carbon dioxide gas diffusion into a thin layer of internal electrolyte. The internal sensing elements are within a durable Delrin® body.
The HI4105 is ideal for a variety of applications including water, beverage, and wine applications.
The HI4105 carbon dioxide ion selective electrode is a complete potentiometric cell that contains both a silver/silver chloride (Ag/AgCl) reference and a pH measurement element. These elements are housed within a plastic body with a chloride ion-containing electrolyte, and are isolated from the sample by a gas permeable membrane made of polytetrafluoroethylene (PTFE).
In solution, carbon dioxide reacts with water to form carbonic acid. Carbonic acid dissociates into bicarbonate and carbonate species. The concentrations of carbon dioxide, carbonic acid, bicarbonate, and carbonate species are pH dependent. As the pH of a solution becomes more alkaline, the equilibrium shifts to favor carbonate as the predominant species. A more acidic pH favors carbon dioxide as the predominant species.
For the HI4105 to measure carbon dioxide accurately, it is important that the pH of the calibration standards and sample be lowered with the carbon dioxide ionic strength adjuster (ISA) solution. The carbon dioxide ISA ensures the pH of solutions is kept between pH 4.2 and 5.2. In this pH range, the carbonate and bicarbonate in solution are converted to carbon dioxide for an accurate measurement. The HI4105 carbon dioxide gas sensing ISE can measure from 400µM (17.6 mg/L) to 0.02M (880 mg/L).
Conversion Table for CO2
Multiply |
|
moles/L (M) to ppm (mg/L) |
4.4 x 104 |
ppm (mg/L) to M (moles/L) |
2.273 x 10-3 |
Replaceable PTFE Membranes - Gas sensing ISEs are combination electrodes that detect dissolved gases in a solution. The sensing element is separated from the sample solution by a gas permeable, PTFE membrane that is easy to replace.
Durable Delrin® Body - The probe body is composed of durable Delrin plastic.
BNC Connection - The HI4105 has a universal BNC connector for easy connection to any benchtop meter with a female BNC probe input.
Type | combination |
---|---|
Measurement Range | Carbon dioxide (CO2) 0.02M to 4•10-4M, 880 to 17.6 mg/L (ppm) |
Optimum pH Range | pH 4.2 to 5.2 |
Temperature Range | 0 to 40 °C |
Approximate Slope | +54 mV |
Diameter | 12 mm |
Overall Length | 120 mm |
Body Material | Delrin |
Cable | coaxial; 1 m (3.3’) |
Connection | BNC |