Measure pH/mV (pH) or mV/Rel. mV (ORP) with Temperature
HI6221 is a streamlined benchtop meter with a large touch screen display, comprised of a housing and an integrated pH / ORP measurement module.
Compact and easy to operate, the benchtop meter is delivered with Hanna Instruments HI1131B double junction combination pH electrode, together with HI7662-TW temperature probe.
HI1131B is a glass body, double junction, refillable pH electrode with an indicating sensor made of High Temperature (HT) glass. The double junction reference and HT glass design allow the HI1131B to be used in a wide variety of applications including samples with metals and elevated temperatures. Probe connection to the unit is secured through a galvanically isolated BNC connection. HI7662-TW stainless steel temperature probe allows the meter to automatically temperature compensate (ATC) pH measurements.
This system responds to a complex range of measurement and monitoring requirements, providing accuracy, reproducibility, and reliability. HI6221 is supplied with an electrode holder that has a flexible arm. The holder can be mounted quickly and provides secure support for electrodes while taking measurements in sample containers.
Customize calibration options such as Last Calibration, Automatic, semi-automtic or manual calibration, First Calibration Point, daily or periodic Calibration Reminder, and buffer Groups.
The benchtop unit has a 7-inch color display with 800 x 480p resolution. The capacitive, multi-touch screen supports video playback and data plotting.
This key will enter the system menu where User accounts, System Settings, and Logging can be configured. The Help menu is also accessed on the system menu screen.
Opens sensor setup parameters.
When Direct/Autohold is selected, measurement reading is held on display when measurement stability is reached. This option removes the subjective nature of stability as a measurement that has not reached equilibrium will not be used.
When not selected, sample measurements are displayed continuously.
SKU | HI6221 |
---|---|
Product Name | Advanced pH/ORP Benchtop Meter- HI6221 |
Quote Required | Yes |
pH Range | -2.0 to 20.0 pH; -2.00 to 20.00 pH; -2.000 to |
pH Range | 20.000 pH |
pH Resolution | 0.1 pH; 0.01 pH; 0.001 pH |
pH Accuracy | ±0.1 pH; ±0.01 pH; ±0.002 pH (±1 last significant digit) |
mV Range | ±2000.0 mV |
mV Resolution | 1 mV; 0.1 mV |
mV Accuracy | ±0.2 mV ±1 last significant digit |
Temperature Range | -20.0 to 120.0 °C; -4.0 to 248.0 °F; 253.0 to 393.0 K |
Temperature Resolution | 0.1 °C / 0.1 °F / 0.1 K |
Temperature Accuracy | ±0.2 °C / ±0.4 °F / ±0.2 K |
Relative mV offset range | ±2000.0 mV |
pH Calibration Calibration points | Up to 5 |
pH Calibration Type | Automatic; Semiautomatic; Manual |
pH Calibration Standard buffers | Hanna and NIST |
pH Calibration Standard buffers | pH 1.68, 3.00, 4.01, 6.86, 7.01, 9.18, 10.01, 12.45'' |
pH Calibration Custom buffers | Up to 5 |
pH Calibration Custom group | Up to 5 |
pH Calibration 1st calibration point | Offset or Points (user setting) |
pH Calibration Reminder | Disabled |
pH Calibration Reminder | Daily |
pH Calibration Reminder | Periodic |
Temperature Compensation | Automatic or Manual |
Reading Modes | Direct; Direct/Autohold |
Reading Stability criteria | Accurate; Medium; Fast |
Reading Isopotential | 7.000 or 4.010 |
Reading Sampling rate | 1000 ms |
pH Views Basic | Measurement (pH, mV, Rel.mV, Abs.mV) |
pH Views Basic | Temperature, Stability status |
pH Views Simple GLP | Basic view information |
pH Views Simple GLP | Last calibration date, electrode offset, average slope, and electrode condition |
pH Views Full GLP | Simple GLP information and calibration point details |
pH Views Table | Measurements updated every second are displayed in table |
pH Views Graph (Plot) | pH (or mV) and temperature versus time graph can be panned or zoomed (pinch-to-zoom technology) |
Logging Type | Automatic, Manual, Autohold |
Logging Number of records | 50 000 maximum per file / Stores at least 1 000 000 data points per user'' |
Logging Automatic interval | 1, 2, 5, 10, 30 seconds |
Logging Automatic interval | 1, 2, 5, 10, 15, 30, 60, 120, 150, 180 minutes'' |
Logging Sample ID | Incremental mode |
Logging Export option | .csv file format |
Logging Export option | Users Up to 9 users and admin. account (default) |
Connectivity USB-A | 2 ports for keyboard input or USB thumb drive |
Connectivity USB-C | 1 port for PC connectivity and USB-C type thumb drive |
Connectivity Wi-Fi & Ethernet | FTP / Web server Log transfer and download Email |
Connectivity RS232 | Connecting peripherals |
Power supply | DC adapter 100-240AC to 24VDC 2.5A |
Environment | 0 - 50 °C / 32 - 122 °F / 273 - 323 K maximum 95% RH non-condensing |
Dimensions | 205 x 160 x 77 mm (8.0 x 6.2 x 3.0 “) |
Weight |
Approximately 1.2 kg (2.65 lbs.) |