详情
HannYa Microcomputer Controlled Semi-Automatic Impact Testing Machine (HY-JBW-300B)
Product Description:
Discover precision and efficiency with the HANNYA HY-JBW-300B Microcomputer Controlled Semi-Automatic Impact Testing Machine. This advanced testing solution falls under the Impact Testing Machine category, delivering unparalleled accuracy and reliability in impact testing procedures.
Designed to streamline testing processes, the HY-JBW-300B boasts microcomputer control and semi-automatic functionality, empowering you to conduct impact tests with utmost precision and minimal manual intervention. Whether you're in research, quality control, or educational settings, this machine offers a seamless experience for analyzing material strength and resilience.
Trust in the HANNYA brand's commitment to excellence, as the HY-JBW-300B embodies the perfect fusion of cutting-edge technology and robust engineering. Elevate your impact testing capabilities and drive insightful results with the HANNYA HY-JBW-300B Microcomputer Controlled Semi-Automatic Impact Testing Machine.
Specifications:
1) Impact energy: 300J、 150J
2) Scale range and division value:
Energy range |
0~300J |
0~150J |
Division value per cell |
2J |
1J |
3) Pendulum moment: |
Pendulum impact energy |
300J |
150J |
Pendulum moment |
160.7695N.m |
80.3848N.m |
4) Pendulum advance angle: 150º
5) Distance from pendulum center to impact point (test center): 750mm
6) Impact velocity: 5.2m/s
7) Span of sample support: 40mm
8) The arc radius at the end of the specimen support: 1~1.5mm
9) The inclination angle of the supporting surface of the specimen support: 0º
10) The arc radius of the impact knife: 2~2.5mm
11) Including angle of impact knife: 30º
12) Thickness of impact knife: 16mm
13) Specimen specification: 10×10×55mm
14) Weight of testing machine: about 450kg
15) External dimensions of the testing machine: 2124×600×1340mm
16) Power supply: three-phase four-wire system 50Hz 380V
17) Motor power: 380V 180W
18) The minimum resolution of computer display: 0.1J
19) Working conditions:
- The room temperature is within the range of 10-35℃;
- The relative humidity is not more than 85%;
- In an environment without corrosive media and strong magnetic field interference;
- Installed on a concrete foundation with a thickness of not less than 300mm or fixed on a foundation greater than 880 kg;
- Adjust the level of the installation reference surface on the machine base to within 0.5/1000.
Manufacturer #Indonesia #Thailand #Philippines #Singapore #Vietnam
更多 Redmark Industry Sdn Bhd 相关资料