The probe is supplied with a calibration certificate.
The probe is interchangeable without calibration to a specific device and regardless of the length of the cable.
MEASURED VALUE
Temperature
Relative humidity
TEMPERATURE SENSOR
Measuring range -10 to +60 °C
Accuracy ±0.4 °C
Response time t90 < 1 min, temperature step 20 °C, air flow 1 m/s
HUMIDITY SENSOR
Measuring range 0 to 100 % RV
Accuracy ±1.8 %RV in the range of 0 to 90 % RH at 23 °C
Response time t90 < 9 s (humidity step 60 %RH, constant temperature)
Hysteresis ±1 % RV
Linearity ±1 % RV
DEW POINT
Measuring range -60 to +60 °C
Accuracy see the manual
GENERAL TECHNICAL DATA
Operating temperature -10 to +60 °C
Protection class IP30
Connector ELKA
Dimensions (without cable) 63 x 16 mm
Cable length 1 m, 2 m, 5 m, 10 m or 15 m
Weight cca 55 g
Warranty 3 years