MEASURED VALUE
Temperature
Relative humidity
Counted values (dew point…)
Two-state signal
TEMPERATURE SENSOR
Measuring range -30 to +105 °C
Accuracy ±0.4 °C
Resolution 0.1 °C
HUMIDITY SENSOR
Measuring range 0 to 100 % RH
Accuracy ±2.5 % RH from 5 to 95 % at 23 °C
Resolution 0.1% RH
DEW POINT
Measuring range -60 to +80 °C
Accuracy ±1.5 °C at ambient temperature T <25 °C and RH >30 %
Resolution 0.1 °C
RELAY OUTPUTS
Quantity 2
Maximal voltage 50 V
Maximal current 2 A
Maximal power 60 VA
BINARY INPUTS
Quantity 3
Signal for binary inputy dry contact, open collector or two-state voltage signal.
Inputs are not galvanically isolated.
Minimum pulse duration on binary input 500 ms
Voltage across open contact < 3,3 V
Low voltage level 0 to +0,5 V
High voltage level +3,0 to +30 V
GENERAL TECHNICAL DATA
Operating temperature -30 to +80 °C
Channels 1x connectable temperature+humidity probe
Acoustic alarm from built-in beeper - switchable
Counted values humidex, dew point, absolute humidity, specific humidity, mixing ratio, specific enthalpy
Output Ethernet
Measuring interval 2 s
Range of humidity sensor temperature compensation all temperature range
Available temperature units degrees Celsius, degrees Fahrenheit
Communication protocol WWW, ModbusTCP, SNMPv1, SOAP, XML
Alarm protocols E-mail (SMTP authentication is supported), SNMP Trap, Syslog
Power 12Vdc, maximum consumption 1W
Protection class IP30 electronics; IP40 sensors
Dimensions 483 x 44 x 45 mm; length/diameter of external probe 88/18 mm; one rack unit 1U
External probe cable length 1 meter
Weight approx. 1090 g
Warranty 3 years