MEASURED VALUE
Temperature
Relative humidity
CO2
TEMPERATURE SENSOR
Measuring range -30 to +80 °C
Accuracy ±0.4 °C
Resolution 0.1 °C
HUMIDITY SENSOR
Measuring range 0 to 100 % RH
Accuracy ±2.5 % RH from 5 to 95 % at 23 °C
Resolution 0.1% RH
DEW POINT
Measuring range -60 to +80 °C
Accuracy ±1.5 °C at ambient temperature T <25 °C and RH >30 %
Resolution 0.1 °C
CO2 SENSOR
Measuring range 0 to 2000 ppm
Accuracy ±(50ppm +2% from reading) at 25°C and 1013hPa
Resolution 1 ppm
RELAY OUTPUTS
Quantity 2
Maximal voltage 50 V
Maximal current 2 A
Maximal power 60 VA
GENERAL TECHNICAL DATA
Operating temperature -30 to +60 °C
Channels internal temperature, humidity and CO2 sensor
Counted values dew point, absolute humidity, specific humidity, mixing ratio, specific enthalpy
Acoustic alarm from built-in beeper - switchable
Power 9-30 Vdc
Protection class IP30 electronics; IP40 sensors
Dimensions 136 x 213 x 45 mm; stem length 75mm
Weight approx. 360 g
Warranty 3 years