TM-3014   AC/DC Clamp Meter


 
This apparatus conforms to safety standard EN 61010

 

Application

It can be used for CURRENT and VOLTAGE measurements on installations of surge voltage category CAT IV 600 V, CAT III up to 1000 voltage between Phase and Earth (fixed installations) and for current measures up to CAT III 1000V/1200A CATIV 600V/1200A .

 

Features

The apparatus can perform the following measurements:

Specifications


CHARACTERISTICS SPECIFICATIONS(23℃±5℃, below 75%RH )
DC Current
Range Resolution Accuracy Overload Protection
660A 0.1A ±(1.5%+5) 1200A rms
( 60 second)
1200A 1A ±(2.0%+5)
 
AC Current
Range Resolution Accuracy
(50~60 Hz)
Overload Protection
660A 0.1A ±(1.5%+10) 1200A rms
( 60 second)
1200A 1A ±(1.5%+5)
 
DC Voltage
Range Resolution Accuracy Input impedance OverloadProtection
6.6V 1mV ±(0.5%rdg+3 dgt) 11M!& DC 1000V 
AC750Vrms
66V 10mV 10M!&
660V 100mV
1000V 1V ±(0.8%rdg+3 dgt)
 
AC Voltage
Range Resolution Accuracy Input impedance OverloadProtection
6.6V 1mV ±(1.2%rdg+3 dgt) 11M!& DC 1000V 
AC750Vrms
66V 10mV 10M!&
660V 100mV
1000V 1V ±(1.5%rdg+4 dgt)
 
AC voltage 
(TRUE RMS: from 10% to 100% of the range)
Range Resolution Accuracy Input impedance Overload Protection
6.6V 1mV ±(1.2%+10)
50Hz~500 Hz
± (1.5%+10)
500Hz~1KHz
11M!& DC 1000V 
AC750Vrms
66V 10mV 10M!&
660V 100mV
750V 1V ±(1.5%+5)
50Hz~500 Hz
±(2.0%+5)500Hz~1KHz
Input Impedance: 10M!&// less than 100Pf.
Crest factor :  >3:1.
 
Resistance
Range Resolution Accuracy Input impedance Overload Protection
660!& 0.1!& ±(1.0rdg+5dgt) About 3.0VDC DC 1000V 
AC750Vrms
6.6K!& 1!& ±(1.0rdg+3dgt) About 0.45VDC
66K!& 10!&
660K!& 100!&
6.6M!& 1K!&
66M!& 10K!& ±(3.0rdg+3dgt)
 
Continuity test
Range Resolution Accuracy Input impedance Overload Protection
0.1!& <35!& About 1.2V DC 1000V 
AC750Vrms
 
Diode test
Range Resolution Max. Open Loop voltage Overload Protection
10mV About3.0VDC DC 1000V 
AC750Vrms
 
Frequency” tests
Range Resolution Accuracy Input impedance Overload Protection
660Hz 0.1Hz ±(0.5%rdg+3 dgt) >1.5VACrms 
<5VACrms
DC 1000V 
AC750Vrms
6.6Hz 1Hz
66KHz 10Hz
100KHz 100Hz
 
Capacitance Test
Range Resolution Accuracy Overload Protection
6.6uF 0.001uF ±(2.0% rdg + 8dgt) DC 1000V AC
750Vrms
66uF 0.01uF
660uF 0.1uF
6.6mF 1uF  
±(5.0% rdg + 20dgt)
66mF 10uF

 


 