Comet T6340 Specifications |
Technical Parameters |
Value |
Output |
RS232 |
Measured Value |
CO2 + Temperature + Relative Humidity |
Construction Type |
Ambient Air |
Design |
Industrial |
Temperature Measuring Range |
-30 to 60 °C |
Relay Output |
No |
Two-State Input |
No |
Lcd Display |
Yes |
PoE |
No |
Range of CO2 concentration measurement |
0 to 2000 ppm |
Accuracy of CO2 concentration measurement |
±(50ppm +2% from reading) at 25°C and 1013hPa |
Optional range of CO2 concentration measurement |
0 to 10000ppm ±(100ppm +5% from reading) at 25°C and 1013hPa |
Relative humidity range |
0 to 100% |
Accuracy of relative humidity measurement |
±2.5% relative humidity from 5 to 95% at 23°C |
Accuracy of temperature output |
±0.4°C |
Available temperature units |
degrees Celsius, Fahrenheit |
Accuracy and range of dew point temperature output - for more details see graphs |
±1.5°C at ambient temperature T<25°C and RH>30%
range -60 to +80 °C |
Accuracy and range of absolute humidity output |
±1.5g/m3 at ambient temperature T < 25°C
range 0 to 400 g/m3 |
Accuracy and range of specific humidity output |
±2g/kg at ambient temperature T < 35°C
range 0 to 550 g/kg |
Accuracy and range of mixing ratio output |
±2g/kg at ambient temperature T < 35°C
range 0 to 995 g/kg |
Accuracy and range of specific enthalpy output |
±3kJ/kg at ambient temperature T < 25°C
range: 0 to 995 kJ/kg |
Measuring interval of CO2 |
15s |
Temperature operating range |
-30 to +60°C |
Communication protocol |
ModBus RTU and Advantech ADAM compatible protocol |
Communication speed |
110 to 115200 Bd |
Filtering ability of sensor cover |
0.025mm - filter with stainless steel mesh |
Protection of the case with electronics |
IP30 electronics, IP40 sensors |
Power |
9-30Vdc, power consumption approximately 1W |
Dimensions |
88.5 x 170 x 39.5 mm (W x H x D), stem length 75 mm |
Weight |
approximately 175g |