All pipettes have been quality tested according to ISO8655-2:2002 with calibration certificate. The quality control involves gravimetric testing of each pipette with distilled water at 22℃.

Features for TopPette


Features for Multi-channel pipette  

8-channel Adjustable Volume Pipettes

Volume Range

Increment

Test Volume

(μl)

Maximume Permissible Systematic error

  (Inaccuracy)

Maximume Permissible Random error

(Inaccuracy)

     

%

μl

%

μl

0.5-10μl

0.1μl

10

1.50%

0.15

1.50%

0.15

5

2.50%

0.125

2.50%

0.125

1

4.00%

0.04

4.00%

0.04

5-50μl

0.5μl

50

1.00%

0.5

0.50%

0.25

25

1.50%

0.375

1.00%

0.25

5

3.00%

0.15

2.00%

0.1

50-300μl

5μl

300

0.70%

2.1

0.25%

0.75

150

1.00%

1.5

0.50%

0.75

50

1.50%

0.75

0.80%

0.4

 

 

12-channel Adjustable Volume Pipettes

Volume Range

Increment

Test Volume

(μl)

Maximume Permissible Systematic error

   (Inaccuracy)

Maximume Permissible Random error

(Inaccuracy)

     

%

μl

%

μl

0.5-10μl

0.1μl

10

1.50%

0.15

1.50%

0.15

5

2.50%

0.125

2.50%

0.125

1

4.00%

0.04

4.00%

0.04

5-50μl

0.5μl

50

1.00%

0.5

0.50%

0.25

25

1.50%

0.375

1.00%

0.25

5

3.00%

0.15

2.00%

0.1

50-300μl

5μl

300

0.70%

2.1

0.25%

0.75

150

1.00%

1.5

0.50%

0.75

50

1.50%

0.75

0.80%

0.4

 

 

Fixed Volume Pipettes

Volume Range

Increment

Test Volume

(μl)

Maximume Permissible Systematic error

   (Inaccuracy)

Maximume Permissible Random error

(Inaccuracy)

     

%

μl

%

μl

5μl

-

5μl

1.3%

0.065

1.2%

0.06

10μl

-

10μl

0.8%

0.08

0.8%

0.08

20μl

-

20μl

0.6%

0.12

0.5%

0.1

25μl

-

25μl

0.5%

0.125

0.3%

0.075

50μl

-

50μl

0.5%

0.25

0.3%

0.15

100μl

-

100μl

0.5%

0.5

0.3%

0.3

200μl

-

200μl

0.4%

0.8

0.2%

0.4

250μl

--

250μl

0.4%

1.0

0.2%

0.5

500μl

-

500μl

0.3%

1.5

0.2%

1.0

1000μl

-

1000μl

0.3%

3.0

0.2%

2.0

2000μl

-

2000μl

0.3%

6.0

0.15%

3.0

5000μl

-

5000μl

0.3%

15

0.15%

7.5
 

Single-channel Adjustable Volume Pipettes

Volume Range

Increment

Test Volume

(μl)

Maximume Permissible Systematic error

   (Inaccuracy)

Maximume Permissible

Random error

   (Inaccuracy)

     

%

μl

%

μl

0.1-2.5μl

0.05μl

2.5

2.50%

0.0625

2.00%

0.05

1.25

3.00%

0.0375

3.00%

0.0375

0.25

12.00%

0.03

6.00%

0.015

0.5-10μl

0.1μl

10

1.00%

0.1

0.80%

0.08

5

1.50%

0.075

1.50%

0.075

1

2.50%

0.025

1.50%

0.015

2-20μl

0.5μl

20

0.90%

0.18

0.40%

0.08

10

1.20%

0.12

1.00%

0.1

2

3.00%

0.06

2.00%

0.04

5-50μl

0.5μl

50

0.60%

0.3

0.30%

0.15

25

0.90%

0.225

0.60%

0.15

5

2.00%

0.1

2.00%

0.1

10-100μl

1μl

100

0.80%

0.8

0.15%

0.15

50

1.00%

0.5

0.40%

0.2

10

3.00%

0.3

1.50%

0.15

20-200μl

1μl

200

0.60%

1.2

0.15%

0.3

100

0.80%

0.8

0.30%

0.3

20

3.00%

0.6

1.00%

0.2

50-200μl

1μl

200

0.60%

1.2

0.15%

0.3

100

0.80%

0.8

0.30%

0.3

50

1.00%

0.5

0.40%

0.2

100-1000μl

5μl

1000

0.60%

6

0.20%

2

500

0.70%

3.5

0.25%

1.25

100

2.00%

2

0.70%

0.7

200-1000μl

5μl

1000

0.60%

6

0.20%

2

500

0.70%

3.5

0.25%

1.25

200

0.90%

1.8

0.30%

0.6

1000-5000μl

50μl

5000

0.50%

25

0.15%

7.5

2500

0.60%

15

0.30%

7.5

1000

0.70%

7

0.30%

3

2-10ml

0.1ml

10ml

0.60%

60

0.20%

20

5ml

1.20%

60

0.30%

15

2ml

3.00%

60

0.60%

12