BK-AutoHS96 Automatic Nucleic Acid Extraction System is a fully automatic high-throughput equipment with automatic sample addition, nucleic acid extraction and PCR system configuration. With magnetic bead extraction reagents, it is suitable for automatic nucleic acid extraction and purification of 1-96 clinical samples of various types. The flexible automatic liquid handling function can accurately complete sample loading and reagent distribution according to requirements. Humanized software design, simple operation, no manual steps, greatly improving work efficiency.
Features:
1. Accurate pipetting, air pressure correction can adapt to extreme environments such as flat ground, plateau, island, etc., to ensure the accuracy of pipetting
2. 96 samples can be processed within 60 minutes, realizing high-throughput processing of samples, saving time and effort
3. Reagent position and PCR plate position, can be refrigerated at 4℃
4. With high-efficiency filter, ultraviolet disinfection and sterilization, and safety door functions, effectively prevent microbial pollution
5. Multi-threaded control and three-module extraction can run three different extraction programs at the same time
6. Intelligent temperature control, over-temperature protection function
7. Preset multiple experimental programs, strong compatibility, suitable for various types of sample graphic guides, visualized operations
8. Nucleic acid products can be allocated to the 2*96 PCR reaction system to flexibly construct a variety of different PCR detection systems
Application:
It is widely used in various fields such as Entry-Exit Inspection and Quarantine Bureau, Disease Control and Prevention Center, Scientific research institution, Laboratory medicine of all size general hospitals, Blood centers, Independent labs.
Technical Parameter:
Model |
BK-AutoHS96 |
Extraction Method |
Magnetic Bead Method |
Working Mode |
Automatic sampling + Nucleic acid extraction + PCR reaction system addition |
Throughput |
1-96, Linear slide type sample rack |
Extraction Volume |
20-1000ul |
Processing Time |
Complete the processing of 96 samples within 60 minutes (related to reagents) |
Magnetic Bead Recovery |
≥98% |
Temp Range |
RT-105℃, Lysis and elution position |
Temp Accuracy |
0.1℃ |
Heating Method |
Dry bath heating |
Heating Speed |
RT-100℃≤6min |
Shaking Function |
Up and down oscillation (1-5 gears adjustable) |
Extraction Position |
6 (96-well deep well plate) |
Robotic arm |
A robotic arm for adding samples and reagents |
Pipetting Channel |
2 Channel |
Liquid Detection |
Pneumatic liquid level detection principle, intelligent detection of blocked needle |
Pipetting Tip |
50ul,200ul,1000ul, Disposable black conductive needle with filter element |
Tip Amount |
2-3 Tips/sample |
Pipetting Accuracy |
10ul, CV≤3.0%, Accuracy≤5.0%, 50ul Tip |
50ul, CV≤2.0%, Accuracy≤2.0%, 1000ul Tip |
|
100ul, CV≤1.5%, Accuracy≤2.0%, 1000ul Tip |
|
Sample Volume |
2-1000ul |
Working Zone |
2 PCR positions with cooling function |
6 Tip positions for three types of Tips |
|
2 Reagent positions (5ml freezing tube rack position with cooling function, one reserved position) |
|
Protective function |
Start up self-test, Power-off protection, High temperature alarm, Over-temperature protection, Tip removal protection |
Disinfection method |
UV lamp (30W*1, 8W*1) |
Illumination Lamp |
10W LED lamp |
Audible Alarm |
Yes (Red and blue blinking) |
Safety Door Design |
With safety lock function, the safety door is opened and the program is suspended |
Display |
10.1inch touch screen, Windows System |
Scanning |
Optional |
Interface |
LAN interface (Bi-direction LIS optional) |
Contamination control |
Built-in air duct and HEPA filter can effectively filter internal aerosols and prevent cross-contamination |
IAP Function |
Firmware can be upgraded online at any time |
Power Supply |
110/220V,50/60Hz |
External Size (W*D*H) |
1420*850*1842mm |
Package Size (WD*H) |
1535*970*1180mm (Main instrument) 1540*970*1160mm(Base) |
Gross Weight |
360kg(Main instrument) ;190kg(Base) |