All Raynova pipettes have been quality tested according to ISO8655-2:2002 with calibration certificate. The quality control involves gravimetric testing of each pipette with distilled water at 22°.
Features for TopPette
-
Lightweight, ergonomic, low force design
-
Digital display clearly reads volume setting
-
The pipettes cover volume range of 0.1μl to 10ml
-
Easy to calibrate and maintain with tool supplied
-
Design helps avoid repetitive strain injuries
-
Calibrated in accordance with ISO8655. Each pipette supplied with individual test certificate
-
The low part is available for autoclaving
Features for multi-channel TopPette
-
8 and 12 channel pipettes are available for standard 96-well plate
-
Dispensing head rotates for optimum pipetting convenience
-
Individual piston and tip cone assemblies allow easy repair and maintenance
-
Compound material tip cone design allows visual seal verification
-
Can be used with universal style pipette tips
Ordering Information
TopPette Single-channel Adjustable Volume Pipettes |
Cat No. |
Volume Range |
Increment |
Test Volume |
Maximum permissible systematic error (Inaccuracy) |
Maximum permissible random error (Imprecision) |
% |
μl |
% |
μl |
711111010000 |
0.1-2.5μl |
0.05μl |
2.5 |
2.50% |
0.0625 |
2.00% |
0.05 |
1.25 |
3.00% |
0.0375 |
3.00% |
0.0375 |
0.25 |
12.00% |
0.03 |
6.00% |
0.015 |
711111040000 |
0.5-10μl |
0.1μl |
10 |
1.00% |
0.1 |
0.80% |
0.08 |
5 |
1.50% |
0.075 |
1.50% |
0.075 |
1 |
2.50% |
0.025 |
1.50% |
0.015 |
For more Information please