详情
Biological safety cabin
Model:MBH-1300A2
External dimensions (W×D×H):1300×802×1480mm
Workspace dimensions (W×D×H):1200×630×700mm
Net weight:240kg
Product feature
Product advantage
- Filter Efficiency and Absorbing Capacity:
1.For standard chemicals: The specific absorbing capacities of filters for isopropyl alcohol, cyclohexane, hydrochloric acid should not be lower than their standard values. Isopropyl alcohol ≥ 500 g; cyclohexane ≥ 750 g; hydrochloric acid ≥ 1620 g.
2.The residual volume of acetone chemicals after adsorption by double-layer molecular filter is less than 0.25 mg/m3, conforming to GBZ/T 160.55. According to JG/T 385-2012, the determination result of sulfur hexafluoride (SF6) concentration controlled by standard fume hood should be lower than 0.5 ppm.
- Face Velocity: 0.4 - 0.6 m/s in line with JG/T 385-2012.
- Working Blowing Rate: 480 m³/h.
1.LED Panel: 3.5-inch LED intelligent display shows temperature inside cabinet, fan rotation rate, service life countdown for filter, and TVOC concentration probes for both filter outlet and indoor air.
2.This product serves for chemical storage and involves laboratory fire safety. Therefore, the manufacturer's products must be underwritten by a large insurer and there must be a insurance policy certificate. The coverage amount for third-party public liability insurance must be ≧ RMB 5 million yuan.
3.Filter Module: Polypropylene (PP material); the standard configuration is 4 blocks of molecular filters with the same specifications;
4.The manufacturer should conduct recycling and environmentally friendly disposal of waste filters so as to prevent them from harming personnel and the environment. The manufacturer shall prove that it or its entrusted unit hold the national industrial waste disposal qualification.