The INFICON Pirani Gauge Enhanced 500 (PGE500) is equipped with the latest digital convection enhanced Pirani technology available on the market.
Due to the physical properties of convection this type of Pirani offers higher accuracy in the measurement range between 100 to 1000 mbar.
The rugged gauge and sensor design in combination with many factory built in features, such as the bright, sharp and clear OLED display with integrated keypad, RS485/RS232digital interface and 4 selectable analog output signals makes the PGE500 a high value/low cost of ownership choice. All these features qualify this gauge for many applications where an economical vacuum measurement from low to high vacuum range is required.
Granville-Phillips® and Mini-Convectron® are registered trademarks of MKS Instruments, Andover, MA
Measurement range N2 | mbar Torr Pa |
1.3×10-4 … 1333 1×10-4 … 1000 1.3×10-2 … 133'000 |
Accuracy (N2) | ||
1.3×10-4 … 1.3×10-3 mbar | 0.1x10-3 mbar resolution | |
1×10-3 … 530 mbar | % of reading | ±10 |
530 … 1333 mbar | % of reading | ±2.5 |
1×10-4 … 1×10-3 Torr | Torr | 0.1 mTorr resolution |
1×10-3 … 400 Torr | % of reading | ±10 |
400 … 1000 Torr | % of reading | ±2.5 |
Repeatability (N2) | % of reading | ±2 |
|
||
Admissible temperature | ||
Operation (ambient) | °C | 0 … +40 |
Storage | °C | -40 … +70 |
Bakeout (electronics removed) | °C | ≤150 |
|
||
Supply voltage | V (dc) | +12 … +28 |
Setpoint relay: Number of setpoints | Number of setpoints | 2 (single-pole double-throw relays (SPDT)); 1A at 30V (dc) resisiv or V (ac) non-inductive |
Materials exposed exposed to vacuum | exposed exposed to vacuum |
gold plated tungsten, 304 & 316 stainless steel, glass, nickel, Teflon® |