The Agilent PVG-500 / 500S(tungsten) / 502S(nickel) series of Pirani gauges provides the ultimate combination of state-of-art technology and ruggedness within a compact design. The PVG-500 series utilizes the most advanced digital Pirani technology combined with a stainless steel sensor design to meet the various needs of the market today.
Measurement range (Air, O2, CO, N2) |
5 x 10-4 to 1000 mbar (3.75 x 10-4 to 750 Torr) | ||||||
Accuracy (N2) | Units | ||||||
% of reading | + 15% | 1 x 10-3 to 100 mbar (1 x 10-3 to 75 Torr) | |||||
+ 50% | 5 x 10-4 to 1 x 10-3 mbar (3.75 x 10-4 to 1 x 10-3 Torr) | ||||||
+ 50% | 100 to 1000 mbar (75 to 750 Torr) | ||||||
Repeatability (Air) | 1 x 10-3 to 100 mbar (1 x 10-3 to 75 Torr) – % of reading +2% | ||||||
Output signal (measurement signal) | Voltage range | Measurement range | |||||
0 to 10.3 V | 1.9 to 10.0 V | ||||||
Voltage vs. pressure | 1.286 V/decade, logarithmic | ||||||
Error signal | 0 to 0.5 V (filament rupture) | ||||||
Minimum loaded impedance | 10 kΩ, short-circuit proof | ||||||
Response time | 80 ms | ||||||
Adjustment | One tactile switch for both ATM and HV adjustment | ||||||
Identification gauge | 27 kΩ, referenced to supply common | ||||||
Setpoint (2) | PVG-500 | PVG-500S, PVG-502S | |||||
Setting range | None | 2 x 10-3 to 500 mbar (1.5 x 10-3 to 375 Torr) | |||||
Hysteresis | 10% above lower threshold % of reading | ||||||
Relay contact | 30 VDC / 0.5 ADC floating | ||||||
Switching time | <20 ms | ||||||
Supply voltage | At gauge | Ripple | Power consumption | ||||
14 to 30 VDC | ≤1 Vpp | ≤1 W | |||||
Electrical connection | FCC 68/RJ45 appliance connector, 8 poles, male | ||||||
Sensor cable | 8 poles plus shielding | ||||||
Cable length | ≤ 100 meter (330 ft) | ||||||
Materials exposed to vacuum | Glass, Ni, NiFe, tungsten (tungsten version) | ||||||
Admissible temperature | Operation | Storage | Vacuum Connection | ||||
5 to 60 °C | –20 to +65 °C | 80 °C |
Outline Drawing
Description | Part Number |
PVG-500 Pirani, KF16 | PVG500KF16 |
PVG-500 Pirani tungsten filament with setpoints, KF16 | PVG500KF16S |
PVG-502 Pirani nickel filament with setpoints, KF16 for corrosive applications | PVG502KF16S |
Accessories |
|
PVG-500 Replacement sensor (tungsten) recommended for most applications | PVG500KF16RS |
PVG-502 Replacement sensor (nickel) recommended for corrosive applications | PVG502KF16RS |