TRIOS OPUS

TRIOS OPUS

Category: TRIOS Available
For more information, visit our official website at hsa-asia.com

Specifications
  • OPUS UV SPECTRAL SENSOR FOR THE ONLINE MEASUREMENT: .
Description

OPUS is the new generation of spectral sensors for online measurement of nitrogen and carbon compounds. Through the analysis of a full spectrum, OPUS is able to deliver reliable readings for NO3-N, NO2-N, organic ingredients (CODeq, BODeq, DOCeq, TOCeq), and a number of other parameters.

OPUS features the new TriOS G2 interface, allowing fast and easy configuration of sensors by using a web browser. Integration into existing process control systems and external data loggers has never been easier.
With the optional battery pack, mobile applications are also feasible. WiFi connectivity allows laptops, tablets or smartphones to be easily used for control without any special application software or app installation.
 
Light source
 
Xenon flash lamp
 
Detector
High-end miniature spectrometer
256 Channels
200 to 360 nm
0.8 nm/pixel
Measurement principle
Attenuation, spectral analysis
Optical path
0,3 mm, 1 mm, 2 mm, 5 mm, 10 mm, 20 mm, 50 mm
 
 
Parameter
Nitrate NO3-N, Nitrite NO2-N, CODeq, BODeq, DOCeq, TOCeq,
TSSeq, KHP, SAC254, COD-SACeq, BOD-SACeq
Measurement accuracy
see measurement ranges next page
Turbidity compensation
Yes
Data logger
~ 2 GB
T100 response time
2 min
Measurement interval
≥ 1 min
 
Housing material
Stainless steel (1.4571/1.4404) or titanium (3.7035)
Dimensions (L x Ø)
470 mm x 48 mm (bei 10 mm path)
~ 18.5“ x 1.9“ (with 10 mm path)
Weight stainless steel
~ 3 kg (bei 10 mm path)
~ 6.6 lbs (with 10 mm path)
Weight titanium
~ 2 kg (bei 10 mm path)
~ 4.4 lbs (with 10 mm path)
 
Interface digital
Ethernet (TCP/IP), RS-232 or RS-485 (Modbus RTU)
Power consumption
≤ 8 W
Power supply
12...24 VDC (± 10 %)
 
 
Maintenance effort
≤ 0.5 h/month (typical)
Calibration/maintenance interval
24 months
System compatibility
Modbus RTU
Warranty
1 year (EU: 2 years)
 
US: 2 years
 
Max. pressure with SubConn
30 bar
~ 435 psig
Max. pressure with fixed cable
3 bar
~ 43.5 psig
Max. pressure in FlowCell
1 bar, 2...4 L/min
~ 14.5 psig at 0.5 to 1.0 gpm
Protection type
IP68
NEMA 6P
 
Sample temperature
+2...+40 °C
~ +36 °F to +104 °F
Ambient temperature
+2...+40 °C
~ +36 °F to +104 °F
Storage temperature
-20...+80 °C
~ -4 °F to +176 °F
Inflow velocity
0,1...10 m/s
~ 0.33 fps to 33 fps


Measurement Range
 
Parameter
Measurement principle
Unit
Factor
 
path length [mm]
 
 
 
 
0.3
1
2
5
10
20
50
absorbance
spectral
AU**
-
0.01...2.2
0.01...2.2
0.01...2.2
0.01...2.2
0.01...2.2
0.01...2.2
0.01...2.2
absorbance
spectral
1/m
-
50...7300
15...2200
7.5...1100
3...440
1.5...220
0.75...110
0.3...44
Nitrate N-NO3
spectral
mg/L
-
1.0...330
0.3...100
0.15...50
0.06...20
0.03...10
0.015...5
0,006...2
Nitrate NO3
spectral
mg/L
-
4.43...1460
1.33...440
0.67...220
0.27...88
0.13...44
0.067...22
0.030...9
Nitrite N-NO2
spectral
mg/L
-
1.7...500
0.5...150
0.25...75
0.1...30
0.05...15
0.025...7.5
0.01...3
Nitrite NO2
spectral
mg/L
-
5.6...1650
1.65...500
0.82...250
0.33...100
0.17...50
0.083...25
0.033...10
DOCeq
spectral
mg/L
-
17...3300
5.0...1000
2.5...500
1.0...200
0.5...100
0.25...50
0.1...20
TOCeq
spectral
mg/L
-
17...3300
5.0...1000
2.5...500
1.0...200
0.5...100
0.25...50
0.1...20
CODeq
spectral
mg/L
-
100...7300***
30...2200***
15...1100***
6.0...440***
3.0...220***
1.5...110***
0.6...44***
BODeq
spectral
mg/L
-
100...7300***
30...2200***
15...1100***
6.0...440***
3.0...220***
1.5...110***
0.6...44***
KHP
spectral
mg/L
-
17...13300
5.0...4000
2.5...2000
1.0...800
0.5...400
0.25...200
0.1...80
SAC254
Single wavelength
1/m
-
50...7300
15...2200
7.5...1100
3.0...440
1.5...220
0.75...110
0.3...44
COD-SACeq****
Single wavelength
mg/L
1.46
75...10600
22...3200
11...1600
4.4...640
2.2...320
1.1...160
0.44...64
BOD-SACeq*****
Single wavelength
mg/L
0.48
24...3500
7.2...1050
3.6...525
1.44...210
0.72...105
0.36...52.5
0.15...21
 
TSSeq*****
Single wavelength
mg/L
2.6
130...4300
40...1300
20...650
8.0...260
4...130
2.0...65
0.8...26
 
           
under laboratory conditions
** absorbance unit
*** depending on the composition of the COD and BOD (sum parameter)
**** based on KHP (Note: 100 mg/L COD-standard-solution corresponds to 85 mg/L KHP)
***** based on SiO2
 
Note:
1 mg/L N-NO3 corresponds to 4.43 mg/L NO3
1 mg/L N-NO2 corresponds to 3.28 mg/L NO2

Versions

More detail about HSA ASIA SDN BHD
HSA ASIA SDN BHD
HSA ASIA SDN BHD Measuring Instruments Selangor | Flow Meter Supply Kuala Lumpur (KL) | Water Analytical Instruments Supplier Malaysia ~ HSA ASIA SDN BHD
Contact Us flagMalaysia