MEASURED VALUE
Temperature
Relative humidity
Counted values (dew point…)
Atmospheric pressure
Two-state signal
TEMPERATURE SENSOR
Measuring range -30 to +80 °C
Accuracy ±0.4 °C
Resolution 0.1 °C
HUMIDITY SENSOR
Measuring range 0 to 100 % RH
Accuracy ±2.5 % RH from 5 to 95 % at 23 °C
Resolution 0.1% RH
DEW POINT
Measuring range -60 to +80 °C
Accuracy ±1.5 °C at ambient temperature T <25 °C and RH >30 %
Resolution 0.1 °C
ATMOSPHERIC PRESSURE SENSOR SENSOR
Measuring range 600 to 1100 hPa
Accuracy ±1.3 hPa at 23 °C from 800 to 1100 hPa
Resolution 0.1 hPa
RELAY OUTPUTS
Quantity 2
Maximal voltage 50 V
Maximal current 2 A
Maximal power 60 VA
BINARY INPUTS
Quantity 3
Signal for binary inputy dry contact, open collector or two-state voltage signal.
Inputs are not galvanically isolated.
Minimum pulse duration on binary input 500 ms
Voltage across open contact < 3,3 V
Low voltage level 0 to +0,5 V
High voltage level +3,0 to +30 V
GENERAL TECHNICAL DATA
Operating temperature -30 to +80 °C
Channels internal sensor of temperature, humidity sensor and atmospheric pressure
Counted values dew point, absolute humidity, specific humidity, mixing ratio, specific enthalpy
Acoustic alarm from built-in beeper - switchable
Output RS485
Range of humidity sensor temperature compensation all temperature range
Available temperature units degrees Celsius, degrees Fahrenheit
Communication protocol ModBus RTU, ARION, Advantech ADAM
Power 9-30 Vdc
Protection class IP54 electronics; IP40 sensors
Dimensions 136 x 213 x 45 mm; stem length 75 mm
Weight approx. 360 g
Warranty 3 years