Portable 100 lpm aerosol hygiene monitoring device for airborne microbial, microbiological, bacterial, mold, fungi and virus analysis
PCE-AS1 is an air sampling meter used to assess microbiological pollution and monitor hygiene. Designed for use in hospitals, clean rooms and food and beverage, cosmetics and pharmaceutical production facilities, the PCE-AS1 Particle Counter meets the latest requirements for devices used for airborne microbial analysis. Optional ISO calibration lab certificate as an accessory.
The PCE-AS1 Particle Counter uses the aerosol impaction principle or Andersen collision principle to assess air quality. The air taken in at the top of the instrument is pulled through 300 micropores and the resulting laminar air flow is directed onto the surface of an agar petri dish. When the sampling cycle is complete, the petri dish is removed and incubated. The bacterial and fungal counts can then be assessed in accordance with USP 797 and CAG-009 guidance.
The collecting head should be sterilized in an autoclave at 121° C for 30 minutes (sterilization with a cloth with 75% alcohol is also possible.) The petri dish should also be sterilized with 75% alcohol.
Number of micropores in the collecting head | 589 |
Hole diameter of the micropores | 0.8 mm / 0.03'' |
Battery | 6400 mA |
Sample throughput | 100 liters per minute (l/min) or 3.53 cubic feet per minute |
Flow rate at measuring head |
0.38 l/min |
Automatic control of the measuring volume | 10 ... 6000 liters |
Dimension of plate | ø 70 ... 90 x 15 mm / ø 2.76 x 3.54 x 0.59 inches |
Data storage | max. 99 parameters and 2000 measurement values |
Dimensions | ø 120 x 300 mm / ø 4.72 x 11.81 inches |
Standards | ISO 14698-1 |
Weight | 2.6 kg / 5.73 lb |