Calibration of humidity chamber

Calibration of humidity chamber

Calibration of a humidity chamber involves verifying and adjusting its performance to ensure it produces accurate and consistent humidity levels. Here's a step-by-step guide:


1. Preparation

  1. Read the Manual: Familiarize yourself with the chamber's operation and calibration instructions provided by the manufacturer.
  2. Gather Required Tools:
    • Hygrometer or reference instrument (calibrated and traceable to national standards).
    • Data logger (optional for continuous monitoring).
    • Calibration salts or a humidity generator (if required).
    • Tools for making adjustments, as per the chamber's specifications.
  3. Stabilize the Environment: Ensure the chamber is in a stable ambient environment free from significant temperature or humidity fluctuations.

2. Perform Calibration

Option A: Using Reference Instrument

  1. Place the Reference Instrument:

    • Position the calibrated hygrometer inside the chamber at the test point.
    • If the chamber is large, place sensors in multiple locations to ensure uniformity.
  2. Set the Humidity:

    • Program the chamber to specific humidity levels (e.g., 30%, 50%, 70%, and 90% relative humidity).
    • Let the chamber stabilize at each set point.
  3. Record the Readings:

    • Compare the chamber's displayed humidity with the reference instrument's reading.
    • Note deviations at each set point.
  4. Repeat for Temperature Variation:

    • If the chamber has temperature control, perform the same tests at different temperatures to verify combined accuracy.

Option B: Using Saturated Salt Solutions

  1. Prepare Salt Solutions:

    • Use salts like sodium chloride (NaCl) or magnesium chloride (MgCl₂) to create known humidity levels in sealed containers.
    • Refer to a salt reference table for expected RH levels at specific temperatures.
  2. Place Solutions Inside:

    • Place the solution container in the chamber and allow stabilization.
  3. Measure Humidity:

    • Compare the chamber's readings with the expected humidity from the salt solution.

3. Adjust the Chamber (if needed)

  • Use the calibration interface or manual adjustments to correct deviations.
  • Follow the manufacturer’s procedure for entering calibration offsets.

4. Verification

  • After adjustment, recheck all set points to confirm the accuracy.
  • Perform a stability test by running the chamber for extended periods and observing deviations over time.

5. Document Calibration

  • Record:
    • The reference instrument details (serial number, last calibration date).
    • The observed deviations and corrections made.
    • Environmental conditions during calibration.
    • The date and technician's name.

6. Schedule Routine Calibration

  • Regularly calibrate the chamber (e.g., annually or as specified by quality standards).

By maintaining proper calibration, your humidity chamber will provide reliable and precise conditions for testing or storage.